Haase A, Matthaei D, Bartkowski R, Dühmke E, Leibfritz D
University of Bremen, F.R.G.
J Comput Assist Tomogr. 1989 Nov-Dec;13(6):1036-40. doi: 10.1097/00004728-198911000-00016.
Snapshot fast low angle shot (FLASH) magnetic resonance (MR) imaging techniques have been developed to enable real time imaging of MR parameters. The method is based on a 64 x 128 FLASH tomogram acquired within less than 200 ms. This work describes snapshot FLASH MR using a single 180 degrees pulse prior to the acquisition of a series of FLASH images. The experiment creates continuous dynamic inversion recovery (IR) T1 contrast in successive images. The total acquisition time of 16 images displaying the IR behavior is less than 4 s. Representative snapshot FLASH IR MR images of the abdomen of healthy rats and of an implanted hepatic tumor are illustrated.
快速低角度激发(FLASH)磁共振(MR)成像技术已被开发用于实现MR参数的实时成像。该方法基于在不到200毫秒内获取的64×128 FLASH断层图像。这项工作描述了在获取一系列FLASH图像之前使用单个180度脉冲的快照FLASH MR。该实验在连续图像中创建了连续的动态反转恢复(IR)T1对比度。显示IR行为的16幅图像的总采集时间不到4秒。展示了健康大鼠腹部和植入肝肿瘤的代表性快照FLASH IR MR图像。