Improved Standardization of the Bio-Rad Platelia Aspergillus Galactomannan Antigen Sandwich Enzyme Immunoassay Using the DS2 (Dynex) Enzyme-Linked Immunosorbent Assay (ELISA) Processing System.

作者信息

Gorton R L, White P L, Bagkeris E, Cotterall D, Desai R, McHugh T, Kibbler C C

机构信息

UK Clinical Mycology Network (UKCMN) Regional Laboratory, Department of Microbiology, Royal Free Hospital, Hampstead, London, United Kingdom

UKCMN Regional Laboratory, Public Health Wales, Microbiology Cardiff, UHW, Cardiff, United Kingdom.

出版信息

J Clin Microbiol. 2015 Jul;53(7):2072-8. doi: 10.1128/JCM.00157-15. Epub 2015 Apr 15.

Abstract

The galactomannan enzyme immunoassay (GM-EIA) is widely utilized for the diagnosis of invasive aspergillosis (IA). There is inconsistent reproducibility of results between centers when the assay is processed manually. Automation of EIAs can reduce variation. This study investigated the semiautomation of the GM-EIA on the DS2 (Dynex) platform in the following three stages: (i) DS2 GM-EIA method validation with experimental samples, (ii) DS2 retesting of case-defined clinical samples, and (iii) a 12-month audit of DS2 GM-EIA performance. In stage i, Bland-Altman analysis demonstrated a reduced variance between optical density index (ODI) values for samples processed on two DS2 platforms (mean difference, -0.02; limits of agreement [LOA], -0.19 to 0.14) compared with the variance between samples processed manually and on a DS2 platform (mean difference, 0.02; LOA, -0.25 to 0.3). In stage ii, 100% (14/14 samples) qualitative agreement was observed for serum samples from patients with IA, with no significant change in the ODI values when samples were processed on the DS2 platform. A significant decrease in ODI values was observed for control serum samples on the DS2 platform (difference, 0.01; P = 0.042). In stage iii, a significant reduction in the frequency of equivocal results, from 5.56% (136/2,443 samples) to 1.56% (15/961 samples), was observed after DS2 automation (difference, 4.0%; 95% confidence interval [CI], 2.7 to 5.2%; P < 0.01), with an equivalent increase in negative results. This study demonstrates that GM-EIA automation may reduce intersite variability. Automation does not have an impact on the repeatability of truly positive results but contributes to a reduction in false-positive (equivocal) GM-EIA results, reducing the need to retest a significant proportion of samples.

摘要

半乳甘露聚糖酶免疫测定法(GM-EIA)被广泛用于侵袭性曲霉病(IA)的诊断。当该测定法采用手工操作时,各中心之间的结果再现性不一致。酶免疫测定法的自动化可以减少变异。本研究在以下三个阶段调查了DS2(Dynex)平台上GM-EIA的半自动化情况:(i)使用实验样本对DS2 GM-EIA方法进行验证,(ii)对病例定义的临床样本进行DS2重新检测,以及(iii)对DS2 GM-EIA性能进行为期12个月的审核。在第一阶段,Bland-Altman分析表明,与手工操作和DS2平台处理的样本之间的差异相比,在两个DS2平台上处理的样本的光密度指数(ODI)值之间的差异有所减小(平均差异为-0.02;一致性界限[LOA]为-0.19至0.1)。在第二阶段,观察到IA患者血清样本的定性一致性为100%(14/14个样本),当样本在DS2平台上处理时,ODI值没有显著变化。在DS2平台上,对照血清样本的ODI值显著降低(差异为0.01;P = 0.042)。在第三阶段,DS2自动化后,可疑结果的频率从5.56%(136/2443个样本)显著降低至1.56%(15/961个样本)(差异为4.0%;95%置信区间[CI]为2.7至5.2%;P < 0.01),阴性结果相应增加。本研究表明,GM-EIA自动化可能会降低不同地点之间的变异性。自动化对真正阳性结果的重复性没有影响,但有助于减少GM-EIA的假阳性(可疑)结果,从而减少对很大一部分样本进行重新检测的必要性。