Ashokkumar Sigamani, Nambiraj Arunai, Sinha Sujit Nath, Yadav Girigesh, Raman Kothanda, Bhushan Manindra, Thiyagarajan Rajesh
Department of Radiation Oncology, Rajiv Gandhi Cancer Institute & Research Centre, New Delhi, India ; School of Advanced Sciences, VIT University, Vellore, India.
School of Advanced Sciences, VIT University, Vellore, India.
Rep Pract Oncol Radiother. 2015 May-Jun;20(3):170-80. doi: 10.1016/j.rpor.2015.02.001. Epub 2015 Mar 1.
To measure and compare the head scatter factor for 7 MV unflattened and 6 MV flattened photon beam using a home-made designed mini phantom.
The head scatter factor (Sc) is one of the important parameters for MU calculation. There are multiple factors that influence the Sc values, like accelerator head, flattening filter, primary and secondary collimators.
A columnar mini phantom was designed as recommended by AAPM Task Group 74 with high and low atomic number material for measurement of head scatter factors at 10 cm and d max dose water equivalent thickness.
The Sc values measured with high-Z are higher than the low-Z mini phantoms observed for both 6MV-FB and 7MV-UFB photon energies. Sc values of 7MV-UFB photon beams were smaller than those of the 6MV-FB photon beams (0.6-2.2% (Primus), 0.2-1.4% (Artiste) and 0.6-3.7% (Clinac iX (2300CD))) for field sizes ranging from 10 cm × 10 cm to 40 cm × 40 cm. The SSD had no influence on head scatter for both flattened and unflattened beams. The presence of wedge filters influences the Sc values. The collimator exchange effects showed that the opening of the upper jaw increases Sc irrespective of FF and FFF.
There were significant differences in Sc values measured for 6MV-FB and unflattened 7MV-UFB photon beams over the range of field sizes from 10 cm × 10 cm to 40 cm × 04 cm. Different results were obtained for measurements performed with low-Z and high-Z mini phantoms.
使用自制的小型体模测量并比较7MV非均整和平坦化6MV光子束的头部散射因子。
头部散射因子(Sc)是计算MU的重要参数之一。有多个因素会影响Sc值,如加速器头部、均整滤过器、初级和次级准直器。
按照美国医学物理学家协会任务组74的建议,设计了一个柱状小型体模,采用高原子序数和低原子序数材料,用于在10cm和dmax剂量水等效厚度处测量头部散射因子。
对于6MV-FB和7MV-UFB光子能量,用高原子序数材料测量的Sc值高于用低原子序数小型体模观察到的值。对于10cm×10cm至40cm×40cm的射野尺寸,7MV-UFB光子束的Sc值小于6MV-FB光子束的Sc值(在Primus直线加速器上为0.6-2.2%,在Artiste直线加速器上为0.2-1.4%,在Clinac iX(2300CD)直线加速器上为0.6-3.7%)。源皮距(SSD)对均整和平坦化光束的头部散射均无影响。楔形滤过器的存在会影响Sc值。准直器交换效应表明,上颌开口会增加Sc值,与均整滤过(FF)和无均整滤过(FFF)无关。
在10cm×10cm至40cm×40cm的射野尺寸范围内,6MV-FB和非均整7MV-UFB光子束测量的Sc值存在显著差异。使用低原子序数和高原子序数小型体模进行测量得到了不同结果。