Suppr超能文献

对南非夸祖鲁-纳塔尔省健康学龄儿童中用于检测贫血的HemoCue设备诊断准确性的评估。

Evaluation of the diagnostic accuracy of the HemoCue device for detecting anaemia in healthy school-aged children in KwaZulu-Natal, South Africa.

作者信息

Gwetu T P, Chhagan M K

机构信息

Department of Public Health, School of Clinical Medicine, College of Health Sciences, Nelson R Mandela School of Medicine, University of KwaZulu-Natal, Durban, South Africa.

出版信息

S Afr Med J. 2015 Sep 21;105(7):596-9. doi: 10.7196/SAMJnew.7919.

Abstract

BACKGROUND

The prevalence of anaemia in school-aged children is reported to be high (>10%), yet neither the onset of anaemia nor the disease causing it is easily established. Any form of anaemia, even if mild, can compromise children's health and survival. This study was conducted to generate data to support or reject use of the HemoCue device as a potential point-of-care method for haemoglobin (Hb) assessment in field and primary healthcare settings.

OBJECTIVE

To assess the validity of the HemoCue in relation to the gold-standard laboratory method.

METHODS

A cross-sectional study of children aged 6-8 years, analysing the diagnostic accuracy of the HemoCue in determining Hb levels in venous blood. Agreement between the HemoCue and laboratory techniques was evaluated using the Bland-Altman plot. The intra-class correlation coefficient was used to assess within-subject variability of measured Hb.

RESULTS

A trend of underestimation of Hb values was noted. The mean Hb with the HemoCue was 11.70 g/dL and that with the laboratory method 12.19 g/dL. The mean difference between the two methods was 0.49 g/dL, with a standard deviation of 0.77 g/dL (95% confidence interval -0.59--0.38). Discrepancies>1 g/dL were identified in 14.1% of cases. Bias increased with increasing Hb values.

CONCLUSION

The HemoCue was found to be comparable to the standard laboratory method for determining Hb concentrations in school-aged children. Its usefulness for screening healthy children was demonstrated, although a full blood count is recommended if anaemia or iron deficiency is suspected.

摘要

背景

据报道,学龄儿童贫血患病率较高(>10%),然而贫血的发病情况及其病因均不易确定。任何形式的贫血,即使是轻度贫血,都会影响儿童的健康和生存。本研究旨在收集数据,以支持或反对将HemoCue设备作为现场和基层医疗环境中血红蛋白(Hb)评估的潜在即时检测方法。

目的

评估HemoCue与金标准实验室方法相比的有效性。

方法

对6至8岁儿童进行横断面研究,分析HemoCue在测定静脉血Hb水平方面的诊断准确性。使用Bland-Altman图评估HemoCue与实验室技术之间的一致性。组内相关系数用于评估测量Hb的受试者内变异性。

结果

发现有低估Hb值的趋势。HemoCue测得的平均Hb为11.70 g/dL,实验室方法测得的平均Hb为12.19 g/dL。两种方法的平均差异为0.49 g/dL,标准差为0.77 g/dL(95%置信区间-0.59--0.38)。在14.1%的病例中发现差异>1 g/dL。偏差随Hb值的增加而增大。

结论

发现HemoCue在测定学龄儿童Hb浓度方面与标准实验室方法相当。证明了其在筛查健康儿童方面的有用性,不过如果怀疑有贫血或缺铁,建议进行全血细胞计数。