The effect of object size on the sensitivity of single photon emission computed tomography: comparison of two CZT cardiac cameras and an Anger scintillation camera.

作者信息

Bienenstock Elazar A, Ennis Marguerite

机构信息

Department of Nuclear Medicine, Etobicoke General Hospital of William Osler Health System, 101 Humber College Blvd., Markham, ON, Canada, M9V 1R8.

Scarborough Cardiac Diagnostic Centre, 2391 Eglinton Ave E, Toronto, ON, Canada, M1K 2M5.

出版信息

EJNMMI Phys. 2014 Dec;1(1):97. doi: 10.1186/s40658-014-0097-5. Epub 2014 Dec 31.

Abstract

BACKGROUND

Heart sizes vary greatly across the spectrum of patients referred for myocardial perfusion imaging. We therefore performed a phantom study to explore under controlled circumstances how count rates change when different volumes containing the same amount of activity are scanned. Two dedicated cadmium-zinc-telluride cameras, the D-SPECT (Spectrum Dynamics, Caesarea, Israel) and Discovery 530c (D530c, GE Healthcare, Haifa, Israel), and the conventional SPECT Anger (A-SPECT, GE Healthcare, Haifa, Israel) camera are included in the study.

METHODS

Different heart sizes were represented by syringes of various column heights mimicking a range of cardiac diameters. Syringes with fixed activity were scanned at five different volumes by successively adding non-radioactive water to the syringes. This procedure was repeated five times on each of the three cameras. Raw count rates were recorded for each scan to determine whether count rates changed with syringe column height.

RESULTS

Using mixed-effect regression modeling, a linear relationship was found between count rate and water column height. For the D-SPECT, D530c, and A-SPECT, the changes in count rate for each centimeter increase in water column height were -1.75, +0.28, and -0.022 kilocounts per min per MBq, respectively (95% confidence intervals -1.89 to -1.61, 0.19 to 0.36, and -0.035 to -0.009); all effects are significantly different from each other and significantly different from zero. Average coefficients of variation were 0.080, 0.028, and 0.009.

CONCLUSIONS

The D-SPECT demonstrated a significant progressive increase in count rate related to decreasing size of the imaged object. D530c count rate increased slightly with increasing column height. The Anger SPECT showed minimally increased count rates with decreasing column height, an order of magnitude smaller than the D-SPECT based on their relative coefficients of variation.

摘要

背景

在接受心肌灌注成像检查的患者中,心脏大小差异很大。因此,我们进行了一项体模研究,以在可控条件下探索当扫描含有相同活度的不同体积时计数率如何变化。该研究纳入了两台专用的碲化镉锌相机,即D-SPECT(Spectrum Dynamics公司,以色列凯撒利亚)和Discovery 530c(D530c,通用电气医疗集团,以色列海法),以及传统的SPECT Anger相机(A-SPECT,通用电气医疗集团,以色列海法)。

方法

用不同柱高的注射器模拟一系列心脏直径,以代表不同的心脏大小。通过依次向注射器中加入非放射性水,在五个不同体积下对装有固定活度的注射器进行扫描。在三台相机中的每一台上,该过程重复五次。记录每次扫描的原始计数率,以确定计数率是否随注射器柱高而变化。

结果

使用混合效应回归模型,发现计数率与水柱高度之间存在线性关系。对于D-SPECT、D530c和A-SPECT,水柱高度每增加1厘米,计数率的变化分别为每分钟每兆贝克勒尔-1.75、+0.28和-0.022千计数(95%置信区间为-1.89至-1.61、0.19至0.36和-0.035至-0.009);所有效应彼此之间均有显著差异,且与零有显著差异。平均变异系数分别为0.080、0.028和0.009。

结论

D-SPECT显示,随着成像对象尺寸减小,计数率显著逐步增加。D530c的计数率随柱高增加略有增加。Anger SPECT显示,随着柱高降低,计数率略有增加,根据其相对变异系数,其增加幅度比D-SPECT小一个数量级。