Suppr超能文献

一种基于多用途全身CzT的新型相机的评估:与双头Anger相机的比较及首批临床图像

Evaluation of a new multipurpose whole-body CzT-based camera: comparison with a dual-head Anger camera and first clinical images.

作者信息

Desmonts Cédric, Bouthiba Mohammed Abdeldjalil, Enilorac Blandine, Nganoa Catherine, Agostini Denis, Aide Nicolas

机构信息

Department of Nuclear Medicine, Caen University Hospital, Avenue de la côte de Nacre, 14033, Caen Cedex 9, France.

Normandie University, Caen, France.

出版信息

EJNMMI Phys. 2020 Mar 17;7(1):18. doi: 10.1186/s40658-020-0284-5.

Abstract

BACKGROUND

Evaluate the physical performance of the VERITON CzT camera (Spectrum Dynamics, Caesarea, Israel) that benefits from new detection architecture enabling whole-body imaging compared to that of a conventional dual-head Anger camera.

METHODS

Different line sources and phantom measurements were performed on each system to evaluate spatial resolution, sensitivity, energy resolution and image quality with acquisition and reconstruction parameters similar to those used in clinical routine. Extrinsic resolution was assessed using Tc capillary sources placed successively in air, in a head and in a body phantom filled with background activity. Spectral acquisitions for various radioelements used in nuclear medicine (Tc, I, Tl, In) were performed to evaluate energy resolution by computing the FWHM of the measured photoelectric peak. Tomographic sensitivity was calculated by recording the total number of counts detected during tomographic acquisition for a set of source geometries representative of different clinical situations. Sensitivity was also evaluated in focus mode for the CzT camera, which consisted of forcing detectors to collect data in a reduced field-of-view. Image quality was assessed with a Jaszczak phantom filled with 350 MBq of Tc and scanned on each system with 30-,20-,10- and 5-min acquisition times.

RESULTS

Extrinsic and tomographic resolution in the brain and body phantoms at the centre of the FOV was estimated at 3.55, 7.72 and 6.66 mm for the CzT system and 2.47, 7.75 and 7.72 mm for the conventional system, respectively. The energy resolution measured at 140 keV was 5.46% versus 9.21% for the Anger camera and was higher in a same manner for all energy peaks tested. Tomographic sensitivity for a point source in air was estimated at 236 counts·s·MBq and increased to 1159 counts·s·MBq using focus mode, which was 1.6 times and 8 times greater than the sensitivity measured on the scintillation camera (144 counts·s·MBq). Head and body measurements also showed higher sensitivity for the CzT camera in particular with focus mode. The Jaszczak phantom showed high image contrast uniformity and a high signal-to-noise ratio on the CzT system, even when decreasing acquisition time by 6-fold. Representative clinical cases are shown to illustrate these results.

CONCLUSION

The CzT camera has a superior sensitivity, higher energy resolution and better image contrast than the conventional SPECT camera, whereas spatial resolution remains similar. Introduction of this new technology may change current practices in nuclear medicine such as decreasing acquisition time and activity injected to patient.

摘要

背景

评估VERITON CzT相机(Spectrum Dynamics公司,以色列凯撒利亚)的物理性能,该相机受益于新型探测架构,与传统双头Anger相机相比,能够进行全身成像。

方法

在每个系统上进行不同的线源和体模测量,以评估空间分辨率、灵敏度、能量分辨率和图像质量,采集和重建参数与临床常规使用的参数相似。使用依次放置在空气中、头部体模和充满本底活性的身体体模中的Tc毛细管源评估外在分辨率。对核医学中使用的各种放射性元素(Tc、I、Tl、In)进行能谱采集,通过计算测量的光电峰的半高宽来评估能量分辨率。通过记录在断层采集期间针对一组代表不同临床情况的源几何形状检测到的总计数来计算断层灵敏度。还在聚焦模式下评估了CzT相机的灵敏度,该模式包括强制探测器在缩小的视野中收集数据。使用填充有350 MBq Tc的Jaszczak体模评估图像质量,并在每个系统上以30分钟、20分钟、10分钟和5分钟的采集时间进行扫描。

结果

对于CzT系统,在视野中心的脑和身体体模中的外在和断层分辨率估计分别为3.55、7.72和6.66 mm,对于传统系统分别为2.47、7.75和7.72 mm。在140 keV处测量的能量分辨率对于Anger相机为5.46%,而对于传统相机为9.21%,并且对于所有测试的能量峰以相同方式更高。空气中点源的断层灵敏度估计为236计数·秒·MBq,使用聚焦模式时增加到1159计数·秒·MBq,分别是在闪烁相机上测量的灵敏度(144计数·秒·MBq)的1.6倍和8倍。头部和身体测量也显示CzT相机具有更高的灵敏度,特别是在聚焦模式下。Jaszczak体模在CzT系统上显示出高图像对比度均匀性和高信噪比,即使采集时间减少6倍时也是如此。展示了代表性临床病例以说明这些结果。

结论

与传统SPECT相机相比,CzT相机具有更高的灵敏度、更高的能量分辨率和更好的图像对比度,而空间分辨率保持相似。这项新技术的引入可能会改变核医学的当前实践,例如减少采集时间和注入患者的活度。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/fb14/7078403/9491e3397c86/40658_2020_284_Fig1_HTML.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验