Diagn Interv Radiol. 2016 Jan-Feb;22(1):101-7. doi: 10.5152/dir.2015.14440.
We aimed to investigate the effect of key imaging parameters on the accuracy of apparent diffusion coefficient (ADC) maps using a phantom model combined with ADC calculation simulation and propose strategies to improve the accuracy of ADC quantification.
Diffusion-weighted imaging (DWI) sequences were acquired on a phantom model using single-shot echo-planar imaging DWI at 1.5 T scanner by varying key imaging parameters including number of averages (NEX), repetition time (TR), echo time (TE), and diffusion preparation pulses. DWI signal simulations were performed for varying TR and TE.
Magnetic resonance diffusion signal and ADC maps were dependent on TR and TE imaging parameters as well as number of diffusion preparation pulses, but not on the NEX. However, the choice of a long TR and short TE could be used to minimize their effects on the resulting DWI sequences and ADC maps.
This study shows that TR and TE imaging parameters affect the diffusion images and ADC maps, but their effect can be minimized by utilizing diffusion preparation pulses. Another key imaging parameter, NEX, is less relevant to DWI and ADC quantification as long as DWI signal-to-noise ratio is above a certain level. Based on the phantom results and data simulations, DWI acquisition protocol can be optimized to obtain accurate ADC maps in routine clinical application for whole body imaging.
我们旨在使用体模模型结合表观扩散系数(ADC)计算模拟,研究关键成像参数对ADC图准确性的影响,并提出提高ADC定量准确性的策略。
在1.5T扫描仪上,通过改变包括平均次数(NEX)、重复时间(TR)、回波时间(TE)和扩散准备脉冲等关键成像参数,使用单次激发回波平面成像DWI在体模模型上采集扩散加权成像(DWI)序列。针对不同的TR和TE进行DWI信号模拟。
磁共振扩散信号和ADC图取决于TR和TE成像参数以及扩散准备脉冲的数量,但不取决于NEX。然而,选择长TR和短TE可用于最小化它们对所得DWI序列和ADC图的影响。
本研究表明,TR和TE成像参数会影响扩散图像和ADC图,但通过利用扩散准备脉冲可将其影响最小化。另一个关键成像参数NEX,只要DWI信噪比高于一定水平,与DWI和ADC定量的相关性就较小。基于体模结果和数据模拟,可优化DWI采集方案,以便在全身成像的常规临床应用中获得准确的ADC图。