[现代呼吸机在顺应性和阻力改变时的容积损失(作者译)]

[The loss of volume in modern respirators during change of compliance and resistance (author's transl)].

作者信息

Ehehalt V, Kluge E, Mottner J

出版信息

Prakt Anaesth. 1977 Apr;12(2):138-44.

PMID:267913
Abstract

Respirators with constant volume proved to have a distinct loss of originally set tidal volume, when mechanical respiratory disturbances are present. In cases with decreased compliance the loss of volume is mainly caused by compression of the basic volume of the apparatus. In cases with increased bronchial flow resistance - unfavourable inspiration - expiration ratio, too short or even lacking endinspiratory pressure plateau result in the distinct decrease of the set respiratory minute volume, which then requires an appropriate correction. The degree of volume loss was measured and analysed in 10 respirators which are in current use.

摘要

当存在机械性呼吸干扰时,定容呼吸机的初始设定潮气量会明显下降。在顺应性降低的情况下,容量损失主要是由设备基本容量的压缩引起的。在支气管气流阻力增加、吸气-呼气比不利、吸气末压力平台过短甚至缺乏的情况下,会导致设定的呼吸分钟通气量明显下降,这就需要进行适当的校正。对10台目前正在使用的呼吸机的容量损失程度进行了测量和分析。