Schlimgen R
Prakt Anaesth. 1977 Dec;12(6):505-10.
A two-compartment lung model with a bronchial obstruction was ventilated with different respirators: Engström E 200, ER 300, ECS 2000, Bennett MA 1 B, Sandoz M 250 and Dräger Servoventilator 900 under same conditions. Redistribution of air, the socalled "pendelluft", was found in all respirators. The amount of "pendelluft" produced by E 200 and M 250 was much more lower than that produced by respirators with low internal compliance, ECS 2000 and Servo 900. Endinspiratory plateau means more even distribution of ventilation volume despite airway obstruction.
在相同条件下,使用不同的呼吸机对带有支气管阻塞的双腔肺模型进行通气:恩斯特龙E 200、ER 300、ECS 2000、贝内特MA 1 B、山德士M 250和德尔格伺服呼吸机900。在所有呼吸机中均发现了空气的再分布,即所谓的“摆动气”。E 200和M 250产生的“摆动气”量远低于内部顺应性低的呼吸机ECS 2000和伺服900产生的量。吸气末平台期意味着尽管存在气道阻塞,但通气量分布更均匀。