Gupta Shalini, Gupta Ashwani K, Verma Minni, Singh Kamaljit, Kaur Amandeep, Chopra Brinder, Kaur Vaneet
Department of Biochemistry, Gian Sagar Medical College and Hospital, Ram Nagar, Patiala District, Punjab, India.
Department of Pharmacology, Gian Sagar Medical College and Hospital, Ram Nagar, Patiala District, Punjab, India.
Int J Appl Basic Med Res. 2016 Jan-Mar;6(1):50-3. doi: 10.4103/2229-516X.174010.
Recent guidelines for estimation of glucose recommend the use of citrate buffer tubes to inhibit glycolysis if the sample cannot be cooled immediately and separated within 30 min. These tubes are currently not available in India. We prepared the citrate tubes and compared the glucose results obtained with sodium fluoride tubes.
Random blood samples of 44 apparently healthy volunteers were collected in three pairs of citrate buffer and sodium fluoride tubes during September to October 2013. They were labeled as 0 h, 1 h and 2 h samples indicating a delay in centrifugation to separate plasma. Glucose was analyzed on the fully auto analyzer in duplicates using glucose oxidase-peroxidase method.
The mean glucose concentrations at 0 h in citrate tubes were 105.8 ± 19.5 mg/dl compared to 99.6 ± 18.3 mg/dl in sodium fluoride tube. There was statistically significant difference in the glucose levels measured in plasma separated from citrate buffer tube and sodium fluoride tube at 0 h, 1 h, and 2 h. The difference between citrate and sodium fluoride tube results ranged from 6.1 mg/dl at 0 h to 7.4 mg/dl at 2 h. Glucose levels decreased significantly at 2 h in both citrate and sodium fluoride tubes.
There is a significant decrease in glucose levels in sodium fluoride tubes even with immediate separation of plasma. There is urgent need to standardize the preanalytical conditions for glucose estimation so that effective inhibition of glycolysis can be done.
近期的血糖估算指南建议,如果样本不能立即冷却并在30分钟内分离,应使用柠檬酸盐缓冲管来抑制糖酵解。目前印度没有这种管子。我们制备了柠檬酸盐管,并比较了用氟化钠管获得的血糖结果。
2013年9月至10月期间,从44名明显健康的志愿者中随机采集血液样本,分别置于三对柠檬酸盐缓冲管和氟化钠管中。它们被标记为0小时、1小时和2小时样本,表明离心分离血浆存在延迟。使用葡萄糖氧化酶-过氧化物酶法在全自动分析仪上对葡萄糖进行双份分析。
柠檬酸盐管中0小时的平均葡萄糖浓度为105.8±19.5mg/dl,而氟化钠管中为99.6±18.3mg/dl。在0小时、1小时和2小时,从柠檬酸盐缓冲管和氟化钠管中分离的血浆中测得的葡萄糖水平存在统计学显著差异。柠檬酸盐管和氟化钠管结果之间的差异范围从0小时的6.1mg/dl到2小时的7.4mg/dl。在柠檬酸盐管和氟化钠管中,2小时时葡萄糖水平均显著下降。
即使立即分离血浆,氟化钠管中的葡萄糖水平也会显著下降。迫切需要规范血糖估算的分析前条件,以便能够有效抑制糖酵解。