七种不同种植体/基台连接设计及其改良后的重复拧紧和松开扭矩的测量:一项体外研究。

Measurements of Repeated Tightening and Loosening Torque of Seven Different Implant/Abutment Connection Designs and Their Modifications: An In Vitro Study.

机构信息

Boston University, Goldman School of Dental Medicine.

出版信息

J Prosthodont. 2018 Feb;27(2):153-161. doi: 10.1111/jopr.12467. Epub 2016 Mar 22.

Abstract

PURPOSE

Repeated tightening and loosening of the abutment screw may alter its mechanical and physical properties affecting the optimal torque and ultimate reliability of an implant/abutment connection. The purpose of this study was to evaluate the effect of repeated tightening and loosening of implant/abutment screws on the loosening torque of implant/abutment connections of commercially available implant systems.

MATERIALS AND METHODS

Seven different implant/abutment connections and their modifications were tested. The screws of each system were tightened according to the manufacturer's specifications. After 20 minutes the screws were loosened. This procedure was repeated ten times, and the differences between the 1st and 10th cycle were expressed as a percentage change RTq(%) and correlated with initial torque, the number of threads, the length of shank, and thread surface area employing Spearman's analysis.

RESULTS

All systems showed significant differences in residual torque (RTq) value (p < 0.05) between the 1st and 10th cycle except groups 6 and 11 (p > 0.05). All connections but group 3 (p = 1.000) showed a significant change from the initial torque (ITq) to the RTq values. The first successive RTq values increased in two connection groups 1 and 2. The remaining connections showed reduced RTq values ranging from -1.2 % (group 5) to -23.5% (group 6). The RTq values declined gradually with every repeated tightening in groups 1, 2, 3, 8, 9, 11, 12. In group 2, after the tenth tightening the RTq was still above the ITq value. Only length of shank demonstrated a correlation with the RTq(%) change over the successive tightening loosening cycles (p < 0.05).

CONCLUSION

Repeated tightening and loosening of implant/abutment screws caused varying torque level changes among the different systems. These observations can probably be attributed to connection design.

CLINICAL IMPLICATIONS

Limiting the number of tightening/loosening cycles in clinical and laboratory procedures is advisable for most of the implant systems tested.

摘要

目的

反复拧紧和松开基台螺丝可能会改变其机械和物理性能,从而影响种植体/基台连接的最佳扭矩和最终可靠性。本研究的目的是评估反复拧紧和松开种植体/基台螺丝对商业上可用种植体系统的种植体/基台连接的松动扭矩的影响。

材料和方法

测试了七种不同的种植体/基台连接及其改良件。每个系统的螺丝都按照制造商的规格拧紧。20 分钟后,螺丝被松开。这个过程重复了十次,第 1 次和第 10 次循环之间的差异用松动扭矩(RTq)的百分比变化 RTq(%)表示,并与初始扭矩、螺纹数量、柄部长度和螺纹表面积用 Spearman 分析相关联。

结果

所有系统在第 1 次和第 10 次循环之间的剩余扭矩(RTq)值均有显著差异(p < 0.05),除了第 6 组和第 11 组(p > 0.05)。除第 3 组(p = 1.000)外,所有连接的初始扭矩(ITq)和 RTq 值均有显著变化。两个连接组 1 和 2 的第一个连续 RTq 值增加。其余的连接显示 RTq 值降低,范围从-1.2%(组 5)到-23.5%(组 6)。在组 1、2、3、8、9、11、12 中,RTq 值随着每一次重复拧紧而逐渐下降。在第 2 组中,第十次拧紧后 RTq 值仍高于 ITq 值。只有柄部长度与连续拧紧/松开循环中的 RTq(%)变化有相关性(p < 0.05)。

结论

反复拧紧和松开种植体/基台螺丝会导致不同系统之间的扭矩水平发生变化。这些观察结果可能归因于连接设计。

临床意义

对于大多数测试的种植体系统,在临床和实验室程序中限制拧紧/松开循环的次数是明智的。