Continuous transcutaneous gas monitoring in adults on regular dialysis treatment.

作者信息

Agazia B, Guarda L, Lombini C, Saporiti E

机构信息

Nephrology and Dialysis Department, Venice City Center, Italy.

出版信息

Int J Artif Organs. 1989 Feb;12(2):103-10.

PMID:2707903
Abstract

We tested a transcutaneous (tc) oxygen (O2) and carbon dioxide (CO2) sensor in 15 normal adults and in 20 patients on regular dialysis treatment (RDT). We compared the tc gas values (ptcO2, PtcCO2) with the gas tension values in arterial blood samples (PaO2, PaCO2), in normal adults and in RDT-patients at the end of a 30-min test and six times during RDT. During the test, PtcO2 correlated with PaO2 both in normal adults (r 0.72 p less than 0.01) and in RDT patients (r 0.41 p less than 0.05). PtcCO2 correlated with PaCO2 (r 0.59 in normal adults and r 0.76 in RDT). During 14 acetate dialyses the changes were expressed as % delta from time 0. % delta PaO2 was -12.7 at 60 min, + 4.8 at 240 min; % delta PtcO2 -6.0 at 60 min, + 9.7 at 240 min; % delta PaO2 -17.7 at 240 min, -1.8 1 h later; % delta PtcCO2 was -15.8 at 240 min, -3.2 1 h later. Both in normal adults and in RDT, patients there was a good relationship between PtcCO2 and PaCO2 values. In normal adults the absolute PtcO2 values were always lower than PaO2 (-14.4 +/- 10 mmHg); in RDT-patients this difference was more pronounced (-27.6 +/- 15.1) and is probably attributable to the lower Hb levels (6.7 +/- 1.1 vs 12.4 +/- 1.2). During RDT the pattern was the same between tc and gas tension values.(ABSTRACT TRUNCATED AT 250 WORDS)