Kleinman B, Powell S
Department of Anesthesiology, Loyola University Stritch School of Medicine, Maywood, IL 60153.
J Clin Monit. 1989 Apr;5(2):111-5. doi: 10.1007/BF01617884.
Methods used in clinical practice to increase the damping of a transducer hydraulically coupled to an intraarterial blood-pressure monitoring system often decrease the undamped natural frequency of the system. This leads to spuriously high systolic and low diastolic pressure readings. The ROSE damping device is being marketed as a possible solution to the problem. We tested the dynamic response of three different catheter systems, with various pressure-tubing lengths of 1 to 7 feet (30.5 to 213.4 cm), with and without the ROSE damping device. The device was able to substantially increase damping and at the same time maintain the undamped natural frequency. Typically it increased the damping coefficient from a minimum of 0.17 +/- 0.01 to a minimum of 0.33 +/- 0.01, while never significantly decreasing the undamped natural frequency. In testing a sample of 25 devices we did observe, however, a wide variability in damping characteristics among different devices. Damping coefficients ranged between 0.19 and 1.20.
临床实践中用于增加与动脉内血压监测系统液压耦合的换能器阻尼的方法,常常会降低系统的无阻尼固有频率。这会导致收缩压读数假性偏高和舒张压读数假性偏低。ROSE阻尼装置作为该问题的一种可能解决方案正在推向市场。我们测试了三种不同导管系统的动态响应,压力管路长度从1至7英尺(30.5至213.4厘米)不等,测试时分别带有和不带有ROSE阻尼装置。该装置能够大幅增加阻尼,同时保持无阻尼固有频率。通常它会将阻尼系数从最低的0.17±0.01增加到最低的0.33±0.01,而且从未显著降低无阻尼固有频率。然而,在测试25个装置的样本时,我们确实观察到不同装置之间的阻尼特性存在很大差异。阻尼系数在0.19至1.20之间。