Freund Kristin A, Kieves Nina R, Hart Juliette L, Foster Sasha A, Jeffery Unity, Duerr Felix M
Am J Vet Res. 2016 Jul;77(7):749-55. doi: 10.2460/ajvr.77.7.749.
OBJECTIVE To evaluate accuracy and reliability of 3 novel goniometers for measurement of canine stifle joint angles and compare the results with those obtained with a universal goniometer (UG). SAMPLE 8 pelvic limbs from 4 canine cadavers. PROCEDURES Each limb was secured to a wooden platform at 3 arbitrarily selected fixed stifle joint angles. Goniometry was performed with 2 smartphone-based applications (novel goniometers A and B), a digital goniometer (novel goniometer C), and a UG; 3 evaluators performed measurements in triplicate for each angle with each device. Results were compared with stifle joint angle measurements on radiographs (used as a gold standard). Accuracy was determined by calculation of bias and total error, coefficients of variation were calculated to estimate reliability, and strength of linear association between radiographic and goniometer measurements was assessed by calculation of correlation coefficients. RESULTS Mean coefficient of variation was lowest for the UG (4.88%), followed by novel goniometers B (7.37%), A (7.57%), and C (12.71%). Correlation with radiographic measurements was highest for the UG (r = 0.97), followed by novel goniometers B (0.93), A (0.90), and C (0.78). Constant bias was present for all devices except novel goniometer B. The UG and novel goniometer A had positive constant bias; novel goniometer C had negative constant bias. Total error at 50° and 100° angles was > 5% for all devices. CONCLUSIONS AND CLINICAL RELEVANCE None of the devices accurately represented radiographically measured stifle joint angles. Additional veterinary studies are indicated prior to the use of novel goniometers in dogs.
目的 评估3种新型测角仪测量犬 stifle 关节角度的准确性和可靠性,并将结果与通用测角仪(UG)获得的结果进行比较。样本 来自4具犬尸体的8个骨盆肢。程序 将每个肢体以3个任意选择的固定 stifle 关节角度固定在木制平台上。使用2种基于智能手机的应用程序(新型测角仪A和B)、数字测角仪(新型测角仪C)和UG进行测角;3名评估人员使用每种设备对每个角度进行3次重复测量。将结果与X线片上的stifle关节角度测量结果(用作金标准)进行比较。通过计算偏差和总误差来确定准确性,计算变异系数以估计可靠性,并通过计算相关系数来评估X线片和测角仪测量之间的线性关联强度。结果 通用测角仪的平均变异系数最低(4.88%),其次是新型测角仪B(7.37%)、A(7.57%)和C(12.71%)。与X线片测量的相关性通用测角仪最高(r = 0.97),其次是新型测角仪B(0.93)、A(0.90)和C(0.78)。除新型测角仪B外,所有设备均存在恒定偏差。通用测角仪和新型测角仪A存在正恒定偏差;新型测角仪C存在负恒定偏差。所有设备在50°和100°角度的总误差均>5%。结论和临床意义 没有一种设备能准确代表X线片测量的stifle关节角度。在犬中使用新型测角仪之前,需要进行更多的兽医研究。