Suppr超能文献

基于连续静脉微透析的葡萄糖监测设备在危重症患者中的点准确性和趋势准确性:一项前瞻性研究。

Point and trend accuracy of a continuous intravenous microdialysis-based glucose-monitoring device in critically ill patients: a prospective study.

作者信息

Leopold J H, van Hooijdonk R T M, Boshuizen M, Winters T, Bos L D, Abu-Hanna A, Hoek A M T, Fischer J C, van Dongen-Lases E C, Schultz M J

机构信息

Department of Intensive Care, Academic Medical Center, Room C3-311, Meibergdreef 9, 1105 AZ, Amsterdam, The Netherlands.

Department of Medical Informatics, Academic Medical Center, Amsterdam, The Netherlands.

出版信息

Ann Intensive Care. 2016 Dec;6(1):68. doi: 10.1186/s13613-016-0171-3. Epub 2016 Jul 19.

Abstract

BACKGROUND

Microdialysis is a well-established technology that can be used for continuous blood glucose monitoring. We determined point and trend accuracy, and reliability of a microdialysis-based continuous blood glucose-monitoring device (EIRUS(®)) in critically ill patients.

METHODS

Prospective study involving patients with an expected intensive care unit stay of ≥48 h. Every 15 min, device readings were compared with blood glucose values measured in arterial blood during blocks of 8 h per day for a maximum of 3 days. The Clarke error grid, Bland-Altman plot, mean absolute relative difference and glucose prediction error analysis were used to express point accuracy and the rate error grid to express trend accuracy. Reliability testing included aspects of the device and the external sensor, and the special central venous catheter (CVC) with a semipermeable membrane for use with this device.

RESULTS

We collected 594 paired values in 12 patients (65 [26-80; 8-97] (median [IQR; total range]) paired values per patient). Point accuracy: 93.6 % of paired values were in zone A of the Clarke error grid, 6.4 % were in zone B; bias was 4.1 mg/dL with an upper limit of agreement of 28.6 mg/dL and a lower level of agreement of -20.5 mg/dL in the Bland-Altman analysis; 93.6 % of the values ≥75 mg/dL were within 20 % of the reference values in the glucose prediction error analysis; the mean absolute relative difference was 7.5 %. Trend accuracy: 96.4 % of the paired values were in zone A, and 3.3 and 0.3 % were in zone B and zone C of the rate error grid. Reliability: out of 16 sensors, 4 had to be replaced prematurely; out of 12 CVCs, two malfunctioned (one after unintentional flushing by unsupervised nurses of the ports connected to the internal microdialysis chamber, causing rupture of the semipermeable membrane; one for an unknown reason). Device start-up time was 58 [56-67] min; availability of real-time data was 100 % of the connection time.

CONCLUSIONS

In this study in critically ill patients who had no hypoglycemic episodes and a limited number of hyperglycemic excursions, point accuracy of the device was moderate to good. Trend accuracy was very good. The device had no downtimes, but 4 out of 16 external sensors and 2 out of 12 CVCs had practical problems.

摘要

背景

微透析是一项成熟的技术,可用于连续血糖监测。我们测定了一种基于微透析的连续血糖监测设备(EIRUS(®))在重症患者中的点准确性、趋势准确性及可靠性。

方法

对预计在重症监护病房停留≥48小时的患者进行前瞻性研究。每天每15分钟将设备读数与动脉血中测量的血糖值进行比较,每次比较8小时,最多持续3天。采用克拉克误差网格、布兰德-奥特曼图、平均绝对相对差异和血糖预测误差分析来表示点准确性,采用速率误差网格来表示趋势准确性。可靠性测试包括设备、外部传感器以及与该设备配套使用的带有半透膜的特殊中心静脉导管(CVC)的相关方面。

结果

我们收集了12例患者的594对配对值(每位患者65[26 - 80;8 - 97](中位数[四分位间距;总范围])对配对值)。点准确性:93.6%的配对值位于克拉克误差网格的A区,6.4%位于B区;在布兰德-奥特曼分析中,偏差为4.1mg/dL,一致性上限为28.6mg/dL,一致性下限为 - 20.5mg/dL;在血糖预测误差分析中,≥75mg/dL的数值中有93.6%在参考值的20%范围内;平均绝对相对差异为7.5%。趋势准确性:96.4%的配对值位于速率误差网格的A区,3.3%和0.3%分别位于B区和C区。可靠性:16个传感器中有4个不得不提前更换;12个CVC中有2个出现故障(一个是由于未受监督的护士意外冲洗了与内部微透析腔相连的端口,导致半透膜破裂;另一个原因不明)。设备启动时间为58[

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/d8ba/4951389/4d5b7708ff53/13613_2016_171_Fig1_HTML.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验