McNeal D R, Baker L L, Symons J T
IEEE Trans Biomed Eng. 1989 Mar;36(3):301-8. doi: 10.1109/10.19851.
Recruitment characteristics of nerve cuff electrodes implanted in four cats for five months were measured. Monopolar, bipolar, and tripolar configurations were considered. Approximately twice the current was required to achieve a given response using the tripolar configuration as compared with monopolar stimulation. Bipolar stimulation also required more current than monopolar stimulation. Using the recruitment data, a number of strategies for modulating muscle tension were considered. It was shown that both pulse amplitude and pulse duration should be software-selectable to achieve adequate control of muscle tension when using either pulse amplitude modulation or pulse duration modulation. When using pulse amplitude modulation, it was found to be desirable to operate at a low pulse duration in the high end of the allowable range for pulse amplitude. For pulse duration modulation, one should operate at a low pulse amplitude in the high end of the allowable range for pulse duration. The effect of pulse amplitude and pulse duration step size on the maximum step change in muscle tension and the linearity of the recruitment curves were examined. The use of logarithmic steps in the modulation parameter was examined and was shown to result in improved controllability and linearity.
对四只猫植入神经袖套电极五个月后的募集特性进行了测量。研究了单极、双极和三极配置。与单极刺激相比,使用三极配置实现给定反应所需的电流大约是单极刺激的两倍。双极刺激所需的电流也比单极刺激多。利用募集数据,考虑了多种调节肌肉张力的策略。结果表明,在使用脉冲幅度调制或脉冲持续时间调制时,脉冲幅度和脉冲持续时间都应可通过软件选择,以实现对肌肉张力的充分控制。当使用脉冲幅度调制时,发现在脉冲幅度允许范围的高端以低脉冲持续时间操作是可取的。对于脉冲持续时间调制,应在脉冲持续时间允许范围的高端以低脉冲幅度操作。研究了脉冲幅度和脉冲持续时间步长对肌肉张力最大步长变化和募集曲线线性的影响。研究了在调制参数中使用对数步长,结果表明这会提高可控性和线性。