Ofstead Cori L, Wetzler Harry P, Eiland John E, Heymann Otis L, Held Sarah B, Shaw Michael J
Ofstead & Associates, Inc, St Paul, MN.
Ofstead & Associates, Inc, St Paul, MN.
Am J Infect Control. 2016 Dec 1;44(12):1675-1677. doi: 10.1016/j.ajic.2016.06.029. Epub 2016 Sep 7.
Researchers evaluated flexible endoscope damage and contamination levels at baseline and 2 months later. Postcleaning test results exceeded benchmarks for all gastroscopes and no colonoscopes. Microbial growth was found in samples from 47% of fully reprocessed endoscopes at baseline and 60% at follow-up. Borescope examinations identified scratches, discoloration, debris, and fluid inside endoscopes. Irregularities changed over time. Study evaluations allowed damaged and contaminated endoscopes to be identified and re-reprocessed or sent for repairs.
研究人员在基线时以及2个月后评估了软性内窥镜的损伤和污染水平。所有胃镜的清洗后测试结果均超过基准,而结肠镜则没有。在基线时,47%的完全重新处理后的内窥镜样本中发现了微生物生长,随访时这一比例为60%。管道镜检查发现内窥镜内部有划痕、变色、碎片和液体。不规则情况随时间变化。研究评估能够识别受损和受污染的内窥镜,并对其进行再次处理或送去维修。