Hawley Jeffrey R, Kalra Prateek, Mo Xiaokui, Raterman Brian, Yee Lisa D, Kolipaka Arunark
Department of Radiology, The Ohio State University Wexner Medical Center Columbus, Ohio, USA.
Department of Surgical Oncology, The Ohio State University Wexner Medical Center Columbus, Ohio, USA.
J Magn Reson Imaging. 2017 May;45(5):1379-1384. doi: 10.1002/jmri.25511. Epub 2016 Oct 25.
Previous studies of breast MR elastography (MRE) evaluated the technique at magnetic field strengths of 1.5 Tesla (T) with the breast in contact with the driver. The aim of this study is to evaluate breast stiffness measurements and their reproducibility using a soft sternal driver at 3T and compare the results with qualitative measures of breast density.
Twenty-two healthy volunteers each underwent two separate breast MRE scans in a 3T MRI. MRE vibrations were introduced into the breasts at 60 Hz using a soft sternal driver and axial slices were collected using a gradient echo MRE sequence. Mean stiffness measurements were calculated for each volunteer as well as a measure of reproducibility using concordance correlation between scans. Mean stiffness values for each volunteer were assessed and related to amounts of fibroglandular tissue (i.e., breast lobules, ducts, and fibrous connective tissue).
The stiffness values were reproducible with a significant P-value < 0.0001 between two scans with concordance correlation of 0.87 and 0.91 for center slice and grouping all slices, respectively. Volunteers with dense breasts (i.e., higher grades of fibroglandular tissue) had mean stiffness values of 0.96 kPa (center slice) and 0.92 kPa (all slices) while those without dense breasts had mean stiffness values of 0.85 kPa (center slice) and 0.83 kPa (all slices) (P ≤ 0.05).
Breast MRE is a reproducible technique at 3T using a soft sternal driver. Dense breasts had significantly higher stiffness measurements compared with nondense breasts.
2 J. MAGN. RESON. IMAGING 2017;45:1379-1384.
以往关于乳腺磁共振弹性成像(MRE)的研究在1.5特斯拉(T)磁场强度下,乳腺与驱动器接触时评估了该技术。本研究的目的是使用3T的软胸骨驱动器评估乳腺硬度测量及其可重复性,并将结果与乳腺密度的定性测量进行比较。
22名健康志愿者在3T磁共振成像(MRI)中分别接受了两次乳腺MRE扫描。使用软胸骨驱动器以60赫兹将MRE振动引入乳腺,并使用梯度回波MRE序列采集轴向切片。计算每名志愿者的平均硬度测量值以及使用扫描间一致性相关性的可重复性测量值。评估每名志愿者的平均硬度值,并将其与纤维腺组织(即乳腺小叶、导管和纤维结缔组织)的量相关联。
两次扫描之间的硬度值具有可重复性,P值显著<0.0001,中心切片和所有切片分组的一致性相关性分别为0.87和0.91。乳腺致密(即纤维腺组织等级较高)的志愿者中心切片的平均硬度值为0.96千帕,所有切片为0.92千帕,而乳腺不致密的志愿者中心切片的平均硬度值为0.85千帕,所有切片为0.83千帕(P≤0.05)。
使用软胸骨驱动器,乳腺MRE在3T时是一种可重复的技术。与非致密乳腺相比,致密乳腺的硬度测量值显著更高。
2 J.Magn.Reson.Imaging 2017;45:1379 - 1384。