Pearce John Graham, Maddess Ted
*MOptom †PhD Eccles Institute for Neuroscience, John Curtin School of Medical Research, Australian National University, Canberra, Australian Capital Territory, Australia (all authors).
Optom Vis Sci. 2017 Mar;94(3):404-410. doi: 10.1097/OPX.0000000000001022.
To determine the inter-visit test-retest variability (TRV) of a spectral domain optical coherence tomograph, the Topcon 3D OCT-2000, in the measurement of optic nerve head topography, peripapillary retinal nerve fiber layer (pRNFL), and macular ganglion cell complex (GCC) parameters in glaucoma patients. We also examine whether TRV with this instrument varies with the extent of glaucomatous damage.
Twenty-four subjects with varying degrees of glaucoma severity provided 41 eyes with usable results for the study. 3D Disc Retinal Nerve Fiber Layer Analysis and Macula V (GCC) scans were repeated 1 week apart, at the same time of day, to determine the inter-visit TRV. TRV was determined using Bland-Altman limits of agreement (LoA) and the resulting coefficients of repeatability (CR).
The overall horizontal and vertical cup/disc ratio CRs were 0.05 and 0.07, respectively. The GCC CR was 2.9 μm. In contrast, average pRNFL TRV expanded with increasing damage, with the LoA being well fitted by ±(34.67 - 0.294(d)), where d is the pRNFL thickness. A more complex model, with constant LoA of ±5.61 μm at d >82 μm, and linearly expanding TRV below that, achieved marginal significance (P < .06).
The repeatability of GCC measurements with this instrument was excellent. The determination of statistically significant change in average pRNFL should take into account average pRNFL thickness.
确定光谱域光学相干断层扫描仪Topcon 3D OCT - 2000在测量青光眼患者视神经乳头地形图、视乳头周围视网膜神经纤维层(pRNFL)和黄斑神经节细胞复合体(GCC)参数时的访间重测变异性(TRV)。我们还研究了该仪器的TRV是否随青光眼损害程度而变化。
24名青光眼严重程度不同的受试者提供了41只可用于该研究的眼睛。3D视盘视网膜神经纤维层分析和黄斑V(GCC)扫描在一天中的同一时间间隔1周重复进行,以确定访间TRV。使用Bland - Altman一致性界限(LoA)和由此产生的重复性系数(CR)来确定TRV。
总体水平和垂直杯盘比的CR分别为0.05和0.07。GCC的CR为2.9μm。相比之下,平均pRNFL的TRV随着损害程度的增加而扩大,LoA由±(34.67 - 0.294(d))很好地拟合,其中d是pRNFL厚度。一个更复杂的模型,在d > 82μm时LoA恒定为±5.61μm,在该值以下TRV呈线性扩大,达到了边缘显著性(P < .06)。
该仪器测量GCC的重复性极佳。在确定平均pRNFL的统计学显著变化时应考虑平均pRNFL厚度。