Reproducibility of Optical Coherence Tomography Angiography Macular and Optic Nerve Head Vascular Density in Glaucoma and Healthy Eyes.

作者信息

Manalastas Patricia I C, Zangwill Linda M, Saunders Luke J, Mansouri Kaweh, Belghith Akram, Suh Min Hee, Yarmohammadi Adeleh, Penteado Rafaella C, Akagi Tadamichi, Shoji Takuhei, Weinreb Robert N

机构信息

*Hamilton Glaucoma Center, Shiley Eye Institute and Department of Ophthalmology University of California, San Diego, La Jolla, CA †Glaucoma Center, Montchoisi Clinic, Swiss Vision Network, Lausanne, Switzerland ‡Department of Ophthalmology, University of Colorado, Denver, CO §Haeundae Paik Hospital, Inje University, Busan, South Korea ∥Department of Ophthalmology and Visual Sciences, Kyoto University Graduate School of Medicine, Kyoto ¶Department of Ophthalmology, Saitama Medical University, Iruma, Saitama, Japan.

出版信息

J Glaucoma. 2017 Oct;26(10):851-859. doi: 10.1097/IJG.0000000000000768.

Abstract

PURPOSE

Optical coherence tomography angiography (OCT-A) is a noninvasive technology that allows visualization of retinal blood vessels. It is important to determine reproducibility of measurements as low precision can impair its diagnostic capabilities. The purpose of this study is to determine intravisit and intervisit reproducibility of optic nerve head (ONH) and macular vessel density measurements with OCT-A.

PATIENTS AND METHODS

Fifteen healthy volunteers and 14 glaucoma patients completed 2 OCT-A (AngioVue; Optovue Inc.) scanning sessions on each of 2 separate days to assess intravisit and intervisit reproducibility. A series of ONH and macula scans were acquired at each session. Vessel density (%), the proportion of vessel area over the total measurement area was calculated. Reproducibility was summarized using coefficients of variation (CV) and intraclass correlation coefficients calculated from variance component models.

RESULTS

In healthy eyes, the CV of intravisit and intervisit global vessel density measures ranged from 1.8% to 3.2% in ONH scans and 2.5% to 9.0% in macular scans. In glaucoma eyes, the CV of intravisit and intervisit global vessel density measures ranged from 2.3% to 4.1% in ONH scans and 3.2% to 7.9% in macular scans. CVs were lower for global than sectorial measures. Global OCT-A ONH intraclass correlation measurements for the retinal nerve fiber layer in healthy eyes were lower (range: 0.65 to 0.85) than in glaucoma eyes (range: 0.89 to 0.94). Scan size did not make large differences in measurement CVs.

CONCLUSIONS

Reproducibility of OCT-A ONH and macula vessel density measurements is good. Moreover, glaucoma patients have sparser vessel density with poorer reproducibility than healthy subjects.

摘要

目的

光学相干断层扫描血管造影(OCT-A)是一种能够实现视网膜血管可视化的非侵入性技术。确定测量的可重复性很重要,因为低精度会损害其诊断能力。本研究的目的是确定使用OCT-A测量视神经乳头(ONH)和黄斑血管密度的日内和日间可重复性。

患者与方法

15名健康志愿者和14名青光眼患者在2个不同日期的每一天都完成了2次OCT-A(AngioVue;Optovue公司)扫描,以评估日内和日间可重复性。每次扫描时采集一系列ONH和黄斑扫描图像。计算血管密度(%),即血管面积占总测量面积的比例。使用变异系数(CV)和根据方差成分模型计算的组内相关系数来总结可重复性。

结果

在健康眼中,ONH扫描的日内和日间总体血管密度测量的CV范围为1.8%至3.2%,黄斑扫描的CV范围为2.5%至9.0%。在青光眼眼中,ONH扫描的日内和日间总体血管密度测量的CV范围为2.3%至4.1%,黄斑扫描的CV范围为3.2%至7.9%。总体测量的CV低于扇形测量。健康眼中视网膜神经纤维层的OCT-A ONH组内相关测量值(范围:0.65至0.85)低于青光眼眼中的测量值(范围:0.89至0.94)。扫描大小对测量CV没有很大差异。

结论

OCT-A测量ONH和黄斑血管密度的可重复性良好。此外,青光眼患者的血管密度比健康受试者更稀疏,可重复性更差。

相似文献

3
OCTA vessel density changes in the macular zone in glaucomatous eyes.青光眼患者眼部黄斑区光学相干断层扫描血管密度的变化
Graefes Arch Clin Exp Ophthalmol. 2018 Aug;256(8):1499-1508. doi: 10.1007/s00417-018-3965-1. Epub 2018 Apr 10.

引用本文的文献

本文引用的文献

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验