使用Stratus光学相干断层扫描技术测量黄斑厚度与信号强度之间的关系。

Relationship between macular thickness measurement and signal strength using Stratus optical coherence tomography.

作者信息

Segal Ori, Shapira Yinon, Gershoni Assaf, Vainer Igor, Nemet Arie Y, Geffen Noa, Mimouni Michael

机构信息

Department of Ophthalmology, Meir Medical Center, Kfar Saba, Israel and the Sackler School of Medicine, Tel Aviv University, Tel Aviv.

Department of Ophthalmology, Rambam Health Care Campus, Haifa, Israel.

出版信息

Clin Ophthalmol. 2016 Nov 11;10:2259-2264. doi: 10.2147/OPTH.S104153. eCollection 2016.

Abstract

PURPOSE

To examine the relationship between signal strength and macular thickness as measured by Stratus optical coherence tomography (OCT)'s fast macular thickness protocol in healthy subjects.

METHODS

In this prospective cross-sectional study 79 eyes of 42 healthy subjects were enrolled. The age, gender, and eye (right vs left) of each subject were recorded. The Stratus OCT fast macular thickness scan protocol was used and the macular thickness was measured with retinal thickness map analysis. Each eye was imaged at least six times to acquire images with signal strengths of 4, 5, 6, 7, 8, and 9 out of 10 via adjustment of the focusing knob. The OCT parameters included in the analysis were thickness in the central 1 mm and in the different quadrants in the 3-mm area.

RESULTS

Overall 79 eyes of 42 patients with a mean age of 38.4±12.4 were included. There was no significant difference between the signal strength measurements obtained with different signal strengths in the central thickness (=0.20). In the superior, nasal, inferior, and temporal quadrants, a signal strength of 8 demonstrated up to 3 µm thicker measurements than a signal strength of 5 (<0.05). In general linear regression analysis, after accounting for age and gender, signal strength did not remain a significant predictor of thickness in any quadrant.

CONCLUSIONS

When using fast map macular measurements, a signal strength of 5 is clinically as efficient as a signal strength of 8 in measuring macular thickness in all quadrants. Insisting on higher signal strength may not be necessary.

摘要

目的

通过Stratus光学相干断层扫描(OCT)的快速黄斑厚度检测方案,研究健康受试者中信号强度与黄斑厚度之间的关系。

方法

在这项前瞻性横断面研究中,纳入了42名健康受试者的79只眼睛。记录每位受试者的年龄、性别和眼睛(右眼与左眼)情况。采用Stratus OCT快速黄斑厚度扫描方案,并通过视网膜厚度图分析测量黄斑厚度。通过调整聚焦旋钮,对每只眼睛至少成像6次,以获取信号强度分别为4、5、6、7、8和9(满分10分)的图像。分析中纳入的OCT参数包括中央1mm区域以及3mm区域内不同象限的厚度。

结果

共纳入42例患者的79只眼睛,平均年龄为38.4±12.4岁。不同信号强度下获得的中央厚度信号强度测量值之间无显著差异(P=0.20)。在上方、鼻侧、下方和颞侧象限,信号强度为8时的测量厚度比信号强度为5时厚达3μm(P<0.05)。在一般线性回归分析中,在考虑年龄和性别因素后,信号强度在任何象限均不再是厚度的显著预测因素。

结论

在使用快速黄斑地图测量时,信号强度为5在测量所有象限的黄斑厚度方面在临床上与信号强度为8一样有效。坚持更高的信号强度可能没有必要。