Van Alstine Anthony W, Viswanathan Suresh
WJB Dorn VAMC, Columbia, SC, USA.
College of Optometry, State University of New York, 33 West 42nd Street, New York, NY, 10036, USA.
Doc Ophthalmol. 2017 Feb;134(1):25-36. doi: 10.1007/s10633-016-9569-3. Epub 2016 Dec 29.
To assess the test-retest reliability of the multifocal photopic negative response (mfPhNR) of normal human subjects.
Multifocal electroretinograms were recorded from one eye of 61 healthy adult subjects on two separate days using a Visual Evoked Response Imaging System software version 4.3 (EDI, San Mateo, California). The visual stimulus delivered on a 75-Hz monitor consisted of seven equal-sized hexagons each subtending 12° of visual angle. The m-step exponent was 9, and the m-sequence was slowed to include at least 30 blank frames after each flash. Only the first slice of the first-order kernel was analyzed. The mfPhNR amplitude was measured at a fixed time in the trough from baseline (BT) as well as at the same fixed time in the trough from the preceding b-wave peak (PT). Additionally, we also analyzed BT normalized either to PT (BT/PT) or to the b-wave amplitude (BT/b-wave). The relative reliability of test-retest differences for each test location was estimated by the Wilcoxon matched-pair signed-rank test and intraclass correlation coefficients (ICC). Absolute test-retest reliability was estimated by Bland-Altman analysis.
The test-retest amplitude differences for neither of the two measurement techniques were statistically significant as determined by Wilcoxon matched-pair signed-rank test. PT measurements showed greater ICC values than BT amplitude measurements for all test locations. For each measurement technique, the ICC value of the macular response was greater than that of the surrounding locations. The mean test-retest difference was close to zero for both techniques at each of the test locations, and while the coefficient of reliability (COR-1.96 times the standard deviation of the test-retest difference) was comparable for the two techniques at each test location when expressed in nanovolts, the %COR (COR normalized to the mean test and retest amplitudes) was superior for PT than BT measurements. The ICC and COR were comparable for the BT/PT and BT/b-wave ratios and were better than the ICC and COR for BT but worse than PT.
mfPhNR amplitude measured at a fixed time in the trough from the preceding b-wave peak (PT) shows greater test-retest reliability when compared to amplitude measurement from baseline (BT) or BT amplitude normalized to either the PT or b-wave amplitudes.
评估正常人类受试者多焦明视负反应(mfPhNR)的重测信度。
使用视觉诱发电位成像系统软件版本4.3(EDI,加利福尼亚州圣马特奥),在两天内从61名健康成年受试者的一只眼睛记录多焦视网膜电图。在75赫兹监视器上呈现的视觉刺激由七个大小相等的六边形组成,每个六边形张角为12°视角。m步指数为9,m序列减慢,以便在每次闪光后至少包含30个空白帧。仅分析一阶核的第一切片。mfPhNR振幅在从基线(BT)开始的波谷中的固定时间以及从前一个b波峰(PT)开始的波谷中的相同固定时间进行测量。此外,我们还分析了归一化到PT(BT/PT)或b波振幅(BT/b波)的BT。通过Wilcoxon配对符号秩检验和组内相关系数(ICC)估计每个测试位置重测差异的相对信度。通过Bland-Altman分析估计绝对重测信度。
根据Wilcoxon配对符号秩检验,两种测量技术的重测振幅差异均无统计学意义。对于所有测试位置,PT测量显示出比BT振幅测量更高的ICC值。对于每种测量技术,黄斑反应的ICC值大于周围位置的ICC值。在每个测试位置,两种技术的平均重测差异均接近零,并且当以纳伏表示时,在每个测试位置两种技术的可靠性系数(COR - 重测差异标准差的1.96倍)相当,但PT测量的%COR(COR归一化到重测和复测振幅的平均值)优于BT测量。BT/PT和BT/b波比率的ICC和COR相当,优于BT的ICC和COR,但不如PT。
与从基线(BT)测量的振幅或归一化到PT或b波振幅的BT振幅相比,从前一个b波峰(PT)开始的波谷中的固定时间测量的mfPhNR振幅显示出更高的重测信度。