Pepe Murad, Kocadal Onur, Iyigun Abdullah, Gunes Zafer, Aksahin Ertugrul, Aktekin Cem Nuri
Ankara Training and Research Hospital, Ankara, Turkey.
Ankara Training and Research Hospital, Ankara, Turkey.
Acta Orthop Traumatol Turc. 2017 Mar;51(2):146-149. doi: 10.1016/j.aott.2016.12.006. Epub 2017 Jan 8.
The aim of our study was to develop a smartphone-aided end vertebra selection method and to investigate its effectiveness in Cobb angle measurement.
Twenty-nine adolescent idiopathic scoliosis patients' pre-operative posteroanterior scoliosis radiographs were used for end vertebra selection and Cobb angle measurement by standard method and smartphone-aided method. Measurements were performed by 7 examiners. The intraclass correlation coefficient was used to analyze selection and measurement reliability. Summary statistics of variance calculations were used to provide 95% prediction limits for the error in Cobb angle measurements. A paired 2-tailed t test was used to analyze end vertebra selection differences.
Mean absolute Cobb angle difference was 3.6° for the manual method and 1.9° for the smartphone-aided method. Both intraobserver and interobserver reliability were found excellent in manual and smartphone set for Cobb angle measurement. Both intraobserver and interobserver reliability were found excellent in manual and smartphone set for end vertebra selection. But reliability values of manual set were lower than smartphone. Two observers selected significantly different end vertebra in their repeated selections for manual method.
Smartphone-aided method for end vertebra selection and Cobb angle measurement showed excellent reliability. We can expect a reduction in measurement error rates with the widespread use of this method in clinical practice.
Level III, Diagnostic study.