[所谓眼压笔眼压计的临床与实验评估]

[Clinical and experimental evaluation of the so-called Tono-Pen tonometer].

作者信息

Draeger J, Rumberger E, Wirt H

机构信息

Augenklinik, Universität, Hamburg.

出版信息

Klin Monbl Augenheilkd. 1989 Sep;195(3):196-202. doi: 10.1055/s-2008-1046439.

Abstract

The measuring accuracy and reliability of the new Tono-Pen tonometer, which works on the Mackay-Marg principle, were tested on isolated human and pigs' bulbi and on 24 patients. On isolated pigs' bulbi the Tono-Pen indicated the true pressure, whereas the pressures measured on isolated human bulbi were too high and the range of scatter was wider. On patients' eyes the values obtained with the Tono-Pen correlated well with those measured with the hand applanation tonometer (r = 0.92), although the values measured with the Tono-Pen again higher (mean 4.5 mm Hg). As indicated by autocorrelation of repeated measurements, the Tono-Pen appears to be mainly responsible for the difference between the two instruments. The standard deviation of the autocorrelation of the Tono-Pen values was 4.5 mm Hg, that of the hand applanation tonometer only 1.2 mm Hg. The considerable standard deviation of the Tono-Pen measurements may be due to the structure of the human cornea, and in particular Bowman's membrane. The reproducibility of Tono-Pen measurements is too low for it to be recommended for ophthalmological diagnosis.

摘要

新型Tono-Pen眼压计基于麦凯-马格原理工作,其测量准确性和可靠性在离体人眼球和猪眼球以及24例患者身上进行了测试。在离体猪眼球上,Tono-Pen眼压计显示出真实压力,而在离体人眼球上测得的压力过高且离散范围更广。在患者眼睛上,Tono-Pen眼压计测得的值与用手持压平眼压计测得的值相关性良好(r = 0.92),尽管用Tono-Pen眼压计测得的值再次偏高(平均高4.5毫米汞柱)。如重复测量的自相关性所示,Tono-Pen眼压计似乎是造成两种仪器测量差异的主要原因。Tono-Pen眼压计测量值的自相关标准差为4.5毫米汞柱,而手持压平眼压计的仅为1.2毫米汞柱。Tono-Pen眼压计测量值的标准差较大可能是由于人角膜的结构,尤其是鲍曼膜。Tono-Pen眼压计测量的可重复性太低,不推荐用于眼科诊断。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验