青光眼患者使用回弹眼压计与压平眼压计(GAT)测量的眼压一致性。

Agreement of patient-measured intraocular pressure using rebound tonometry with Goldmann applanation tonometry (GAT) in glaucoma patients.

机构信息

Department of Ophthalmology and Visual Sciences, The Chinese University of Hong Kong, Hong Kong.

Department of Ophthalmology, Chinese PLA General Hospital, Beijing, China.

出版信息

Sci Rep. 2017 Feb 6;7:42067. doi: 10.1038/srep42067.

Abstract

This study aims to determine the agreement of patient-measured intraocular pressure (IOP) using rebound tonometry with ophthalmologist-measured IOP using Goldmann applanation tonometry (GAT). Fifty-three glaucoma patients used rebound tonometry (Icare ONE, Icare Finland Oy., Finland) to measure their own IOP in ambient environments for 1 week, 5 times per day. Clinic IOP measurements were performed by ophthalmologists using GAT and by patients using rebound tonometry on examination days 1, 4 and 7 of the same week. The agreement between the two tonometries was evaluated by modified Bland-Altman plots and intra-class correlation coefficient (ICC) was determined. Differences in ICCs of them among the three examination days were evaluated by bootstrap resampling analysis. Respective within-measurement ICC of GAT and rebound tonometry were 0.98 and 0.94 on Day 1, 0.98 and 0.93 on Day 4, and 0.96 and 0.92 on Day 7. In a modified Bland-Altman plot, the mean difference ±1 standard deviation (SD) between the two tonometries was 0.15 ± 0.65 mmHg (p = 0.682). Between-measurement ICC were 0.66, 0.76 and 0.73 on the 3 examination days. There was no significant difference among ICCs. In conclusion, patient-measured IOP using rebound tonometry and ophthalmologist-measured IOP using GAT demonstrate good agreement.

摘要

本研究旨在确定使用回弹眼压计(Rebound tonometry)测量的患者眼压(IOP)与使用 Goldmann 压平眼压计(Goldmann applanation tonometry,GAT)测量的眼科医生眼压之间的一致性。53 名青光眼患者在一周内的日常环境中,每天使用回弹眼压计(Icare ONE,Icare Finland Oy.,芬兰)测量 5 次眼压。在同一周的第 1、4 和 7 天,由眼科医生使用 GAT 以及患者使用回弹眼压计进行临床眼压测量。通过改良的 Bland-Altman 图和组内相关系数(intra-class correlation coefficient,ICC)评估两种眼压计之间的一致性,并通过 bootstrap 重采样分析评估它们在 3 个检查日之间的 ICC 差异。在第 1 天、第 4 天和第 7 天,GAT 和回弹眼压计的各自内测量 ICC 分别为 0.98 和 0.94、0.98 和 0.93 以及 0.96 和 0.92。在改良的 Bland-Altman 图中,两种眼压计之间的平均差值±1 标准差(standard deviation,SD)为 0.15±0.65mmHg(p=0.682)。在 3 个检查日之间,组间 ICC 分别为 0.66、0.76 和 0.73。ICC 之间没有显著差异。总之,使用回弹眼压计测量的患者眼压与使用 GAT 测量的眼科医生眼压之间具有良好的一致性。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/54de/5292696/7a09454043f1/srep42067-f1.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验