Erdogan Hakika, Akingol Ziya, Cam Ozlem, Sencan Sadik
Department of Ophthalmology, Faculty of Medicine, Maltepe University, Maltepe, Istanbul, Turkey,
Clin Ophthalmol. 2018 Oct 29;12:2183-2188. doi: 10.2147/OPTH.S177870. eCollection 2018.
The aim of the study was to statistically compare intraocular pressure (IOP) values measured using noncontact tonometer (NCT), Goldmann applanation tonometry (GAT) with fluorescein (fGAT), and GAT without fluorescein (nGAT). The study was also performed to test whether the values obtained using each technique change in accordance with the central corneal thickness (CCT) and refractive and keratometric values.
This study was a prospective study of 188 eyes of 94 healthy volunteers. IOP was measured using fGAT, nGAT and NCT. CCT, refractive values, and keratometric values were measured, and the correlations and differences in the IOP for each tonometer were investigated.
The mean IOP values obtained with the NCT, nGAT, and fGAT were 17.5±3.7, 12.3±2.7, and 12.5±2 mmHg. The mean CCT was 538.2±34.4 µm, the mean refractive value was 0.9±1.2 D, and the mean keratometric value was 43.5±1.5 D. NCT was positively correlated with fGAT and GAT values and was significantly higher than both the values. There were no differences between fGAT and GAT values. No correlation was observed between the CCT and keratometric and refractive values and the difference between NCT and nGAT or fGAT.
Differences in the measurements obtained using nGAT and fGAT were insignificant (>0.05). Both values were positively correlated with NCT measurements ( =0.354, <0.05) and were independent of CCT, keratometry, and refraction values. nGAT appears to be suitable for use in routine clinic practice.
本研究旨在对使用非接触眼压计(NCT)、荧光素眼底血管造影(fGAT)的戈德曼压平眼压计(GAT)以及无荧光素的GAT(nGAT)测量的眼压(IOP)值进行统计学比较。本研究还旨在测试使用每种技术获得的值是否会根据中央角膜厚度(CCT)、屈光和角膜曲率值而变化。
本研究是一项对94名健康志愿者的188只眼睛进行的前瞻性研究。使用fGAT、nGAT和NCT测量眼压。测量CCT、屈光值和角膜曲率值,并研究每种眼压计的IOP相关性和差异。
NCT、nGAT和fGAT获得的平均IOP值分别为17.5±3.7、12.3±2.7和12.5±2 mmHg。平均CCT为538.2±34.4 µm,平均屈光值为0.9±1.2 D,平均角膜曲率值为43.5±1.5 D。NCT与fGAT和GAT值呈正相关,且显著高于两者的值。fGAT和GAT值之间没有差异。未观察到CCT与角膜曲率和屈光值以及NCT与nGAT或fGAT之间的差异存在相关性。
使用nGAT和fGAT获得的测量值差异不显著(>0.05)。这两个值均与NCT测量值呈正相关(=0.354,<0.05),且与CCT、角膜曲率测量和屈光值无关。nGAT似乎适用于常规临床实践。