Otten Volker, Maguire Gerald Q, Noz Marilyn E, Zeleznik Michael P, Nilsson Kjell G, Olivecrona Henrik
Department of Surgical and Perioperative Sciences, Umeå University, Umeå, Sweden.
School of Information and Communication Technology, KTH Royal Institute of Technology, Kista, Sweden.
Biomed Res Int. 2017;2017:3681458. doi: 10.1155/2017/3681458. Epub 2017 Jan 24.
As part of the 14-year follow-up of a prospectively randomized radiostereometry (RSA) study on uncemented cup fixation, two pairs of stereo radiographs and a CT scan of 46 hips were compared. Tantalum beads, inserted during the primary operation, were detected in the CT volume and the stereo radiographs and used to produce datasets of 3D coordinates. The limit of agreement between the combined CT and RSA datasets was calculated in the same way as the precision of the double RSA examination. The precision of RSA corresponding to the 99% confidence interval was 1.36°, 1.36°, and 0.60° for -, -, and -rotation and 0.40, 0.17, and 0.37 mm for -, -, and -translation. The limit of agreement between CT and RSA was 1.51°, 2.17°, and 1.05° for rotation and 0.59, 0.56, and 0.74 mm for translation. The differences between CT and RSA are close to the described normal 99% confidence interval for precision in RSA: 0.3° to 2° for rotation and 0.15 to 0.6 mm for translation. We conclude that measurements using CT and RSA are comparable and that CT can be used for migration studies for longitudinal evaluations of patients with RSA markers.
作为一项关于非骨水泥髋臼杯固定的前瞻性随机放射性立体测量(RSA)研究14年随访的一部分,对46例髋关节的两组立体X线片和一次CT扫描进行了比较。在初次手术期间植入的钽珠在CT容积和立体X线片中被检测到,并用于生成三维坐标数据集。联合CT和RSA数据集之间的一致性界限以与双重RSA检查精度相同的方式计算。RSA在99%置信区间对应的精度,对于 -、 - 和 - 旋转分别为1.36°、1.36°和0.60°,对于 -、 - 和 - 平移分别为0.40、0.17和0.37毫米。CT和RSA之间的一致性界限,对于旋转分别为1.51°、2.17°和1.05°,对于平移分别为0.59、0.56和0.74毫米。CT和RSA之间的差异接近所描述的RSA精度的正常99%置信区间:旋转为0.3°至2°,平移为0.15至0.6毫米。我们得出结论,使用CT和RSA进行的测量具有可比性,并且CT可用于对具有RSA标记物的患者进行纵向评估的迁移研究。