Suppr超能文献

人工耳蜗植入受者激活一天后阻抗场遥测技术的演变

Evolution of impedance field telemetry after one day of activation in cochlear implant recipients.

作者信息

Hu Hao-Chun, Chen Joshua Kuang-Chao, Tsai Chia-Mi, Chen Hsing-Yi, Tung Tao-Hsin, Li Lieber Po-Hung

机构信息

Department of Otolaryngology, Cheng Hsin General Hospital, Taipei, Taiwan.

Institute of Clinical Medicine, National Yang-Ming University, Taipei, Taiwan.

出版信息

PLoS One. 2017 Mar 6;12(3):e0173367. doi: 10.1371/journal.pone.0173367. eCollection 2017.

Abstract

OBJECTIVES

Changes in impedance between 24 hours and one month after cochlear implantation have never been explored due to the inability to switch on within one day. This study examined the effect of early activation (within 24 hours) on the evolution of electrode impedance with the aim of providing information on the tissue-to-electrode interface when electrical stimulation was commenced one day post implantation.

METHODS

We performed a retrospective review at a single institution. Patients who received a Nucleus 24RECA implant system (Cochlear, Sydney, Australia) and underwent initial switch-on within 24 hours postoperatively were included. Impedance measurements were obtained intraoperatively and postoperatively at 1 day, 1 week, 4 weeks, and 8 weeks.

RESULTS

A significant drop in impedance was noted 1 day after an initial activation within 24 hours followed by a significant rise in impedance in all channels until 1 week, after which the impedance behaved differently in different segments. Basal and mid-portion electrodes revealed a slight increase while apical electrodes showed a slight decrease in impedance from 1 week to 8 weeks postoperatively. Impedance was relatively stable 4 weeks after surgery.

CONCLUSIONS

This is the first study to report the evolution of impedance in all channels between initial mapping 1 day and 1 month after cochlear implantation. The underlying mechanism for the differences in behavior between different segments of the electrode may be associated with the combined effect of dynamics among the interplay of cell cover formation, electrical stimulation, and fibrotic reaction.

摘要

目的

由于无法在一天内开机,此前从未探索过人工耳蜗植入术后24小时至1个月期间阻抗的变化情况。本研究检测了早期开机(24小时内)对电极阻抗变化的影响,旨在为植入术后一天开始电刺激时组织与电极界面的情况提供信息。

方法

我们在单一机构进行了一项回顾性研究。纳入接受Nucleus 24RECA植入系统(澳大利亚悉尼科利耳公司)且术后24小时内首次开机的患者。术中及术后1天、1周、4周和8周进行阻抗测量。

结果

在24小时内首次开机后1天,阻抗显著下降,随后所有通道的阻抗在1周前显著上升,此后不同节段的阻抗变化各异。术后1周至8周,基底和中部电极的阻抗略有增加,而顶端电极的阻抗略有下降。术后4周阻抗相对稳定。

结论

这是第一项报告人工耳蜗植入术后1天初次调机至1个月期间所有通道阻抗变化情况的研究。电极不同节段行为差异的潜在机制可能与细胞覆盖形成、电刺激和纤维化反应相互作用中的动态综合效应有关。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/de37/5338792/0cc74eec94db/pone.0173367.g002.jpg