Bahrami Bobak, Ewe Shaun Y P, Hong Thomas, Zhu Meidong, Ong Germane, Luo Kehui, Chang Andrew
Ophthalmic Surg Lasers Imaging Retina. 2017 Apr 1;48(4):319-325. doi: 10.3928/23258160-20170329-06.
To evaluate the repeatability, reliability, and comparability of macular thickness measurements between three optical coherence tomography (OCT) machines in healthy eyes, eyes with diabetic macular edema (DME), and eyes with neovascular age-related macular degeneration (nAMD).
Twenty-three eyes with DME, 26 eyes with nAMD, and 24 healthy eyes as controls were evaluated. Scans were performed using the swept-source Triton (Topcon, Tokyo, Japan), the spectral-domain Cirrus (Carl Zeiss Meditec, Dublin, CA), and the Spectralis (Heidelberg Engineering, Heidelberg, Germany) machines. Scans were evaluated for central macular thickness (CMT), presence of segmentation and fixation imaging artifacts (IA), re-scan reliability, and agreement between machines and groups.
Mean CMT was significantly different between all OCT machines in all groups (P < .01 for all comparisons). Manually correcting IA did not alter these results. There was good scan repeatability among healthy and DME eyes for each machine, but poor repeatability among the nAMD group with the Spectralis (P = .038). IA were significantly increased in the presence of pathology.
There is poor agreement of CMT measurement between OCT machines in healthy eyes and those with DME and nAMD. DME and nAMD have a significant effect on the rate of IA in scans. Care is required when interpreting measurements from different OCT devices in clinical practice and research settings. [Ophthalmic Surg Lasers Imaging Retina. 2017;48:319-325.].
评估三种光学相干断层扫描(OCT)设备对健康眼、糖尿病性黄斑水肿(DME)眼和新生血管性年龄相关性黄斑变性(nAMD)眼黄斑厚度测量的可重复性、可靠性和可比性。
评估了23只DME眼、26只nAMD眼以及24只健康眼作为对照。使用扫频源Triton(日本东京拓普康公司)、谱域Cirrus(美国加利福尼亚州都柏林卡尔蔡司医疗技术公司)和Spectralis(德国海德堡海德堡工程公司)设备进行扫描。对扫描结果评估中心黄斑厚度(CMT)、分割和注视成像伪像(IA)的存在情况、重复扫描可靠性以及设备和组间的一致性。
所有组中所有OCT设备之间的平均CMT均存在显著差异(所有比较P <.01)。手动校正IA并未改变这些结果。每种设备在健康眼和DME眼中扫描具有良好的可重复性,但Spectralis在nAMD组中可重复性较差(P = 0.038)。在存在病变时IA显著增加。
健康眼以及DME和nAMD眼中,OCT设备之间CMT测量的一致性较差。DME和nAMD对扫描中IA的发生率有显著影响。在临床实践和研究环境中解释来自不同OCT设备的测量结果时需要谨慎。[《眼科手术、激光与视网膜成像》。2017;48:319 - 325。]