Otchere Heinz, Sorbara Luigina
School of Optometry & Vision Science, Centre for Contact Lens Research, University of Waterloo, 200 University Avenue West, Waterloo, Ontario N2L 3G1, Canada.
Cont Lens Anterior Eye. 2017 Aug;40(4):217-223. doi: 10.1016/j.clae.2017.05.002. Epub 2017 May 10.
To assess the repeatability and reproducibility of Visante™ OCT and Oculus Pentacam HR® in measuring central (CCT) and topographic corneal thickness (TCT) along the principle meridians in keratoconus (KC) corneae.
Twenty participants diagnosed with KC were recruited. There were two study visits. On the first study visit, two repeated corneal thickness measurements were obtained with each instrument. Measurements were repeated at least 48hrs later in the same order. TCT were recorded in the 90, 180, 45 and 135 meridians at 1mm intervals across 8mm chord.
Mean CCT for Visante™ OCT was 484.97±43.14μm (range: 484.84-486.09) and Oculus Pentacam HR® was 478.86±45.31μm (range: 477.20-480.53). No significant difference in TCT between the two visits (p=0.54) and measurements (p=0.63) for Visante™ OCT. For Oculus Pentacam HR®, no significant difference was found for each visit (p=0.18) but differences existed in the measurements outside of the central region (p=0.001). Tukey post-hoc analysis shows the differences (p˂0.05) were found in the +1 and +4 (supero-temporal) locations in the 135 meridian. Significant differences were found comparing the two instruments, (p<0.05). Bland Altman plots were used to demonstrate the differences between the two instruments and indicate their limits of agreement.
Both instruments gave repeatable measurements as no significant differences were found in most locations in all meridians. Comparing the two instruments, they were not reproducible in all locations.
评估Visante™光学相干断层扫描(OCT)和Oculus Pentacam HR®在测量圆锥角膜(KC)角膜中央角膜厚度(CCT)和沿主要子午线的地形图角膜厚度(TCT)方面的可重复性和再现性。
招募了20名被诊断为KC的参与者。进行了两次研究访视。在第一次研究访视中,使用每种仪器获得了两次重复的角膜厚度测量值。至少48小时后以相同顺序重复测量。在90、180、45和135子午线处,以1mm的间隔在8mm弦长上记录TCT。
Visante™ OCT的平均CCT为484.97±43.14μm(范围:484.84 - 486.09),Oculus Pentacam HR®的平均CCT为478.86±45.31μm(范围:477.20 - 480.53)。Visante™ OCT两次访视之间(p = 0.54)和测量值之间(p = 0.63)的TCT无显著差异。对于Oculus Pentacam HR®,每次访视之间无显著差异(p = 0.18),但在中央区域以外的测量值存在差异(p = 0.001)。Tukey事后分析显示在135子午线的+1和+4(颞上)位置存在差异(p˂0.05)。比较两种仪器时发现显著差异(p<0.05)。使用Bland Altman图来展示两种仪器之间的差异并表明它们的一致性界限。
两种仪器在所有子午线的大多数位置均未发现显著差异,因此测量具有可重复性。比较两种仪器,它们在所有位置并非都具有再现性。