We aimed to determine optimal window settings for poly-energetic (PolyE) and virtual mono-energetic images (MonoE) derived from abdominal angiographic studies on a novel dual-layer spectral detector CT (SDCT) system.
From 50 patients, SDCT datasets PolyE and MonoE at 70 and 40 keV levels were reconstructed and best individual window width and level (BI-W/L) manually assessed. Through regression analysis, the so-called optimized individual (OI-W/L) values were obtained. Subjective image quality parameters and vessel diameters were measured to determine influences of different W/L settings.
Image noise was lower and attenuation and contrast-to-noise ratio were higher in MonoE compared to PolyE (all p ≤ 0.002). Mean BI-W/L values for PolyE, 70, and 40 keV were 637/284, 647/291, and 1568/691, respectively. Mean OI-W/L values were 631/276, 628/286, and 1516/667, respectively. Compared to standard settings, all adjusted W/L settings varied significantly and yielded higher subjective scoring. No between-group differences were found between manually adjusted and mathematically calculated W/L settings.
PolyE and MonoE from abdominal angiographic SDCT studies require appropriate W/L settings especially at low energy reconstruction levels. Individual adjustment reaches the best image quality but is time consuming. From our data, predefined W/L settings of 640/280 (PolyE/MonoE 70 keV) and 1570/690 (MonoE 40 keV) as a non-individualized starting point for abdominal angiographic studies from the novel SDCT system are suggested.
我们旨在确定新型双层能谱探测器 CT(SDCT)系统腹部血管造影研究中多能量(PolyE)和虚拟单能量(MonoE)图像的最佳窗宽和窗位设置。
从 50 名患者中,重建 SDCT 数据集的 PolyE 和 MonoE 在 70keV 和 40keV 水平,并手动评估最佳个体窗宽和窗位(BI-W/L)。通过回归分析,得到所谓的优化个体(OI-W/L)值。测量主观图像质量参数和血管直径,以确定不同 W/L 设置的影响。
与 PolyE 相比,MonoE 的图像噪声更低,衰减和对比噪声比更高(均 p≤0.002)。PolyE、70keV 和 40keV 的平均 BI-W/L 值分别为 637/284、647/291 和 1568/691。平均 OI-W/L 值分别为 631/276、628/286 和 1516/667。与标准设置相比,所有调整后的 W/L 设置均有显著差异,主观评分更高。手动调整和数学计算的 W/L 设置之间没有组间差异。
腹部血管造影 SDCT 研究中的 PolyE 和 MonoE 需要适当的 W/L 设置,尤其是在低能量重建水平下。个体调整可获得最佳图像质量,但耗时。根据我们的数据,建议将 640/280(PolyE/MonoE 70keV)和 1570/690(MonoE 40keV)作为新型 SDCT 系统腹部血管造影研究的非个体化起点的预设 W/L 设置。