Zhou Lian-Juan, Xua Hong-Zhen, Xu Mei-Fang, Hu Yan, Lou Xiao-Fang
Department of Nursing, Children's Hospital, Zhejiang University School of Medicine, No.3333 Bing-sheng Road, Hangzhou, 310051, P.R.China.
Open Med (Wars). 2017 May 22;12:125-130. doi: 10.1515/med-2017-0019. eCollection 2017.
To explore the clinical application of the intracavitary electrocardiogram (IC-ECG) guided Peripherally Inserted Central Catheter (PICC) tip placement among neonates.
the ECGs of neonates are difficult to perform and their wave shapes are of doubtful accuracy due to various interfering factors.
115 neonates were admitted to perform PICC guided by IC-ECG. Logistic regression was performed to analyze all possible influencing factors of the accuracy from the tip placement. The puncture site of the PICC, gestational age, height, weight, basal P/R amplitude and positioning P/R amplitude might be related to the accuracy of IC-ECG location.
The accuracy in the lower extremity was higher than that in the upper extremity. Multivariate logistic regression analysis showed that the weight (Odds Ratio (OR)=1.93, 95%Confidence Interval(CI):1.06-3.50) and positioning P/R amplitude (OR=32.33, 95%CI: 2.02-517.41) are statistically significant risks to the accuracy PICC tip placement.
Possible methods to improve the accuracy might be Catheterizing through lower extremity, keeping the neonates calm, enhancing the electrocardiogram signal and strengthening technical training. Therefore it is practical to perfrom a tip placement by the dynamic change in the P waves from an electrocardiogram (ECG) guided PICC among neonates and as reliable as using X-rays.
探讨腔内心电图(IC-ECG)引导下新生儿经外周静脉穿刺中心静脉置管(PICC)尖端定位的临床应用。
由于各种干扰因素,新生儿心电图难以实施,其波形准确性存疑。
115例新生儿接受IC-ECG引导下的PICC置管。进行逻辑回归分析以确定尖端定位准确性的所有可能影响因素。PICC的穿刺部位、胎龄、身高、体重、基础P/R振幅和定位P/R振幅可能与IC-ECG定位的准确性有关。
下肢置管的准确性高于上肢。多因素逻辑回归分析显示,体重(比值比(OR)=1.93,95%置信区间(CI):1.06 - 3.50)和定位P/R振幅(OR = 32.33,95%CI:2.02 - 517.41)是PICC尖端定位准确性的统计学显著风险因素。
提高准确性的可能方法包括经下肢置管、保持新生儿安静、增强心电图信号和加强技术培训。因此,通过心电图(ECG)引导的PICC中P波的动态变化进行尖端定位在新生儿中是可行的,并且与使用X射线一样可靠。