非侵入性和侵入性即时检测方法与血红蛋白测量参考方法的比较。

Comparison of noninvasive and invasive point-of-care testing methods with reference method for hemoglobin measurement.

作者信息

Avcioglu Gamze, Nural Cemil, Yilmaz Fatma Meriç, Baran Pervin, Erel Özcan, Yilmaz Gülsen

机构信息

Department of Medical Biochemistry, Ankara Yıldırım Beyazıt University Faculty of Medicine, Ankara, Turkey.

Department of Clinical Biochemistry, Ankara Atatürk Training and Research Hospital, Ankara, Turkey.

出版信息

J Clin Lab Anal. 2018 Mar;32(3). doi: 10.1002/jcla.22309. Epub 2017 Aug 23.

Abstract

BACKGROUND

Rapid and practical point-of-care testing (POCT) devices become more popular, especially in blood donation centers for determining predonation hemoglobin (Hb) concentrations. The purpose of this study was to evaluate accordance between the POCT methods and the venous method as the reference to Hb screening.

METHODS

A total of 353 subjects with no known significant health problems were included in the study. Hb screening was performed by two different POCT methods, a noninvasive method (Haemospect, MBR, Germany) and an invasive method (HemoControl, EKF Diagnostic, Germany), and a venous method as the reference (Sysmex XE-2100, Sysmex Europe, Germany). The obtained results were compared.

RESULTS

The sensitivity and the specificity values of the invasive POCT method (83.3%, 87.9%) were higher than the noninvasive POCT method (66.7%, 77.1%). The Bland-Altman analysis was evaluated for both sexes and the bias of the noninvasive POCT method of the males (-0.97 g/dL) was higher than the bias of the invasive POCT method of the males (-0.07 g/dL). We found a better correlation between the invasive POCT method (r = .908) compared with the venous method than the noninvasive POCT method (r = .634).

CONCLUSION

Predonation Hb measurements must be performed with accurate, precise, and practical methods. Although the noninvasive POCT method was practical and painless, it had lower levels of specificity and sensitivity, and more false deferral and pass rates than the invasive POCT method. The POCT methods agreeable to the venous method as the reference might be suitable for Hb screening especially for centers of excessive numbers of blood donation.

摘要

背景

快速实用的即时检验(POCT)设备越来越受欢迎,尤其是在献血中心用于测定献血前血红蛋白(Hb)浓度。本研究的目的是评估POCT方法与作为Hb筛查参考的静脉血检测方法之间的一致性。

方法

本研究共纳入353名无已知重大健康问题的受试者。采用两种不同的POCT方法进行Hb筛查,一种是非侵入性方法(Haemospect,德国MBR公司)和一种侵入性方法(HemoControl,德国EKF诊断公司),并以静脉血检测方法(Sysmex XE - 2100,德国Sysmex欧洲公司)作为参考。对获得的结果进行比较。

结果

侵入性POCT方法的敏感性和特异性值(83.3%,87.9%)高于非侵入性POCT方法(66.7%,77.1%)。对男女分别进行了Bland - Altman分析,男性非侵入性POCT方法的偏差(-0.97 g/dL)高于男性侵入性POCT方法的偏差(-0.07 g/dL)。我们发现,与静脉血检测方法相比,侵入性POCT方法(r = 0.908)的相关性优于非侵入性POCT方法(r = 0.634)。

结论

献血前Hb测量必须采用准确、精确且实用的方法。虽然非侵入性POCT方法实用且无痛,但与侵入性POCT方法相比,其特异性和敏感性较低,假延期和通过率更高。与作为参考的静脉血检测方法一致的POCT方法可能适用于Hb筛查,特别是对于献血量过多的中心。

相似文献

4
5
Point-of-Care Testing Effectiveness on Blood Donor Hemoglobin Testing.
Lab Med. 2021 Jul 1;52(4):364-368. doi: 10.1093/labmed/lmaa102.
6
Evaluation of various methods of point-of-care testing of haemoglobin concentration in blood donors.
Blood Transfus. 2015 Apr;13(2):233-9. doi: 10.2450/2014.0085-14. Epub 2014 Oct 29.
7
Comparison of three noninvasive methods for hemoglobin screening of blood donors.
Transfusion. 2015 Feb;55(2):379-87. doi: 10.1111/trf.12819. Epub 2014 Aug 15.
10
Evaluation of point-of-care testing device for anemia detection: A cross-sectional method comparison study from Thailand.
J Clin Lab Anal. 2021 Oct;35(10):e23976. doi: 10.1002/jcla.23976. Epub 2021 Aug 24.

引用本文的文献

1
The impact of sensor application site on continuous noninvasive haemoglobin trends.
Comput Struct Biotechnol J. 2025 Jul 12;28:249-258. doi: 10.1016/j.csbj.2025.07.013. eCollection 2025.
2
REASONS FOR DEFERRAL IN DEFERRED VOLUNTARY BLOOD DONORS IN EASTERN CROATIA.
Acta Clin Croat. 2023 Apr;62(1):93-105. doi: 10.20471/acc.2023.62.01.12.
4
Non-invasive hemoglobin estimation for preoperative anemia screening.
Transfusion. 2023 Feb;63(2):315-322. doi: 10.1111/trf.17237. Epub 2023 Jan 5.
5
Current Status of Measurement Accuracy for Total Hemoglobin Concentration in the Clinical Context.
Biosensors (Basel). 2022 Dec 8;12(12):1147. doi: 10.3390/bios12121147.
6
Emerging point-of-care technologies for anemia detection.
Lab Chip. 2021 May 18;21(10):1843-1865. doi: 10.1039/d0lc01235a.
8
Individual differences in haemoglobin concentration influence bold fMRI functional connectivity and its correlation with cognition.
Neuroimage. 2020 Nov 1;221:117196. doi: 10.1016/j.neuroimage.2020.117196. Epub 2020 Jul 25.
9
Metabolism-Driven High-Throughput Cancer Identification with GLUT5-Specific Molecular Probes.
Biosensors (Basel). 2018 Apr 10;8(2):39. doi: 10.3390/bios8020039.

本文引用的文献

1
Evaluation of various methods of point-of-care testing of haemoglobin concentration in blood donors.
Blood Transfus. 2015 Apr;13(2):233-9. doi: 10.2450/2014.0085-14. Epub 2014 Oct 29.
2
Comparison of three noninvasive methods for hemoglobin screening of blood donors.
Transfusion. 2015 Feb;55(2):379-87. doi: 10.1111/trf.12819. Epub 2014 Aug 15.
4
Accuracy of noninvasive hemoglobin and invasive point-of-care hemoglobin testing compared with a laboratory analyzer.
Int J Lab Hematol. 2014 Feb;36(1):56-61. doi: 10.1111/ijlh.12118. Epub 2013 Jun 27.
7
Hemoglobin determination with paired emitter detector diode.
Anal Bioanal Chem. 2011 Mar;399(9):3293-7. doi: 10.1007/s00216-010-4358-4. Epub 2010 Nov 3.
8
Capillary and venous haemoglobin levels in blood donors: a 42-month study of 36,258 paired samples.
Vox Sang. 2010 May;98(4):547-53. doi: 10.1111/j.1423-0410.2009.01285.x. Epub 2009 Nov 24.
9
Evaluation of four rapid methods for hemoglobin screening of whole blood donors in mobile collection settings.
Transfus Apher Sci. 2007 Jun;36(3):235-42. doi: 10.1016/j.transci.2007.01.010. Epub 2007 Jun 6.
10
Managing donors and iron deficiency.
Vox Sang. 2004 Jul;87 Suppl 2:22-4. doi: 10.1111/j.1741-6892.2004.00448.x.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验