Eur J Radiol. 2017 Oct;95:124-129. doi: 10.1016/j.ejrad.2017.08.005. Epub 2017 Aug 7.
This meta-analysis aimed to evaluate the accuracy of dual-energy CT (DECT) virtual non-calcium (VNC) imaging for the detection of bone marrow edema (BME).
A systematic literature search up to March 2017 was performed to find relevant original studies. Two reviewers independently selected studies, assessed literature quality, and extracted data. Pooled sensitivity, specificity, area under receiver operating characteristic (AUROC) curve, and other measures of DECT accuracy for detecting BME were calculated using random effects models. Risk of heterogeneity was assessed for the appropriateness of meta-analysis.
Fourteen studies involving 2205 regions of vertebrae, hips, knees, and ankles were included. To evaluate the accuracy of BME detection using DECT, calculations were performed to obtain a pooled sensitivity of 0.812 (95% confidence interval [CI], 0.780-0.841) and specificity of 0.951 (95% CI, 0.940-0.960). The AUROC value was 0.9635. The major potential cause of heterogeneity was bone position. No significant publication bias was present.
DECT VNC imaging gives very good diagnostic performance for BME detection and will likely be an important and common modality for acute assessment in the future.
本荟萃分析旨在评估双能 CT(DECT)虚拟非钙(VNC)成像检测骨髓水肿(BME)的准确性。
系统检索截至 2017 年 3 月的相关原始研究。两名审查员独立选择研究、评估文献质量并提取数据。使用随机效应模型计算用于检测 BME 的 DECT 准确性的汇总敏感性、特异性、受试者工作特征(ROC)曲线下面积(AUROC)和其他度量。评估异质性的风险以确定荟萃分析的适宜性。
纳入了 14 项涉及脊柱、髋关节、膝关节和踝关节的 2205 个区域的研究。为了评估使用 DECT 检测 BME 的准确性,计算出汇总敏感性为 0.812(95%置信区间[CI],0.780-0.841),特异性为 0.951(95% CI,0.940-0.960)。AUROC 值为 0.9635。主要的异质性潜在原因是骨位置。不存在显著的发表偏倚。
DECT VNC 成像对 BME 检测具有很好的诊断性能,可能成为未来急性评估的重要且常见方式。