使用 320 排 CT 获得的虚拟单色图像质量:一项评估迭代重建和体型影响的体模研究。

Image quality of virtual monochromatic images obtained using 320-detector row CT: A phantom study evaluating the effects of iterative reconstruction and body size.

机构信息

Department of Radiology, Keio University School of Medicine, 35 Shinanomachi, Shinjuku-ku, Tokyo, 160-8582, Japan.

Research Park, Keio University School of Medicine, 35 Shinanomachi, Shinjuku-ku, Tokyo, 160-8582, Japan.

出版信息

Eur J Radiol. 2017 Oct;95:212-221. doi: 10.1016/j.ejrad.2017.08.016. Epub 2017 Aug 19.

Abstract

OBJECTIVES

To compare the image quality between virtual monochromatic spectral (VMS) images obtained using 320-row detector CT and polychromatic 120-kVp images reconstructed with or without iterative reconstruction using various phantom sizes.

MATERIALS AND METHODS

Torso phantoms simulating three patient sizes and containing four syringes filled with water or different contrast media (5, 10, 15mgI/mL15mgI/ml) were used. The phantoms were scanned using dual-energy (80/135-kVp) and single-energy (120-kVp) protocols at different settings (20mGy, 12mGy, and 6mGy). VMS images were generated at 1-keV intervals (range, 35-135keV). Both the VMS images and the single-energy 120-kVp images were reconstructed using filtered back projection (FBP) and adaptive iterative dose reduction 3D (AIDR-3D). The signal-to-noise ratio (SNR), and the contrast-to-noise ratio (CNR) were assessed.

RESULTS

Using FBP reconstruction, the SNR and CNR of the VMS images were lower than or similar to those of the 120-kVp images for most dose settings. Using AIDR-3D reconstruction, however, the 70-keV VMS images had higher SNRs and CNRs than the 120-kVp images at most settings.

CONCLUSIONS

The image quality of VMS images with FBP reconstruction tended to be lower than that of the 120-kVp images. With the use of AIDR-3D, however, approximately 70-keV VMS images had a higher image quality than the 120-kVp images.

摘要

目的

比较使用 320 排探测器 CT 获得的虚拟单色光谱(VMS)图像与使用不同大小体模进行多谱 120kVp 图像重建时使用迭代重建的图像质量。

材料与方法

使用模拟三种患者体型的体模,每个体模内有四个注射器,分别装有水或不同对比剂(5、10、15mgI/ml)。在不同设置(20mGy、12mGy 和 6mGy)下使用双能(80/135kVp)和单能(120kVp)协议进行扫描。VMS 图像在 1keV 间隔(范围为 35-135keV)生成。VMS 图像和单能 120kVp 图像均使用滤波反投影(FBP)和自适应迭代剂量降低 3D(AIDR-3D)重建。评估信噪比(SNR)和对比噪声比(CNR)。

结果

使用 FBP 重建,对于大多数剂量设置,VMS 图像的 SNR 和 CNR 低于或与 120kVp 图像相似。然而,使用 AIDR-3D 重建,在大多数设置下,70keV 的 VMS 图像的 SNR 和 CNR 均高于 120kVp 图像。

结论

使用 FBP 重建的 VMS 图像的图像质量往往低于 120kVp 图像。然而,使用 AIDR-3D,大约 70keV 的 VMS 图像的图像质量高于 120kVp 图像。