早期青光眼性眼病中央视野缺损的检测:Humphrey视野计与Octopus视野计的比较
Detection of central visual field defects in early glaucomatous eyes: Comparison of Humphrey and Octopus perimetry.
作者信息
Roberti Gloria, Manni Gianluca, Riva Ivano, Holló Gabor, Quaranta Luciano, Agnifili Luca, Figus Michele, Giammaria Sara, Rastelli Davide, Oddone Francesco
机构信息
IRCSS-Fondazione GB Bietti, Rome, Italy.
DSCMT, University of Rome Tor Vergata, Rome, Italy.
出版信息
PLoS One. 2017 Oct 27;12(10):e0186793. doi: 10.1371/journal.pone.0186793. eCollection 2017.
PURPOSE
To compare the detection rate of central visual field defect (CVFD) between the 30-degree Octopus G1 program (Dynamic strategy) and the HFA 10-2 SITA-Standard test in early glaucoma eyes not showing any CVFD on the HFA 24-2 SITA-Standard test.
METHODS
One eye of 41 early glaucoma patients without CVFD in the central 10° on HFA 24-2 test was tested with both the HFA 10-2 test and the Octopus G1 program 15 minutes apart, in random order. The primary outcome measure was the comparison of CVFD detection rates. Secondary outcome measures comprised the agreement in detecting CVFD, and the comparison of test durations and the numbers of depressed test points outside the central 10-degree area between the HFA 24-2 test and the Octopus G1 program.
RESULTS
The mean age of the population was 65.2±10.1 years, and the mean deviation with HFA 24-2 was -3.26±2.6 dB. The mean test duration was not significantly different between the tests (p = 0.13). A CVFD was present in 33 (80.4%) HFA 10-2 test and in 23 (56.0%) Octopus G1 tests (p = 0.002). The overall agreement between the HFA 10-2 and Octopus G1 examinations in classifying eyes as having or not having CVFD was moderate (Cohen's kappa 0.47). The Octopus G1 program showed 69.6% sensitivity and 100% specificity to detect CVFD in eyes where the HFA 10-2 test revealed a CVFD. The number of depressed test points (p<5%) outside the central 10° area detected with the Octopus G1 program (19.68±10.6) was significantly higher than that detected with the HFA 24-2 program (11.95±5.5, p<0.001).
CONCLUSION
Both HFA 10-2 and Octopus G1programs showed CVFD not present at HFA 24-2 test although the agreement was moderate. The use of a single Octopus G1 examination may represent a practical compromise for the assessment of both central and peripheral visual field up to 30° eccentricity without any additional testing and increasing the total investigation time.
目的
比较30度Octopus G1程序(动态策略)与HFA 10-2 SITA标准测试在早期青光眼眼中中央视野缺损(CVFD)的检测率,这些早期青光眼眼在HFA 24-2 SITA标准测试中未显示任何CVFD。
方法
对41例在HFA 24-2测试中中央10°无CVFD的早期青光眼患者的一只眼睛,分别用HFA 10-2测试和Octopus G1程序进行测试,两次测试间隔15分钟,测试顺序随机。主要观察指标是CVFD检测率的比较。次要观察指标包括CVFD检测的一致性,以及HFA 24-2测试与Octopus G1程序之间测试持续时间的比较和中央10度区域以外压低测试点数量的比较。
结果
研究人群的平均年龄为65.2±10.1岁,HFA 24-2的平均偏差为-3.26±2.6 dB。两次测试的平均测试持续时间无显著差异(p = 0.13)。HFA 10-2测试中有33例(80.4%)存在CVFD,Octopus G1测试中有23例(56.0%)存在CVFD(p = 0.002)。HFA 10-2与Octopus G1检查在将眼睛分类为有或无CVFD方面的总体一致性为中等(Cohen's kappa 0.47)。在HFA 10-2测试显示有CVFD的眼中,Octopus G1程序检测CVFD的敏感性为69.6%,特异性为100%。Octopus G1程序检测到的中央10°区域以外压低测试点数量(19.68±10.6)显著高于HFA 24-2程序检测到的数量(11.95±5.5,p<0.001)。
结论
HFA 10-2和Octopus G1程序均显示在HFA 24-2测试中不存在的CVFD,尽管一致性为中等。使用单次Octopus G1检查可能是一种实际的折衷方案,可用于评估中央和周边视野直至30°偏心度,无需任何额外测试且不增加总检查时间。