高血压患者晨测、餐前测和睡前测家庭血压的可靠性。

Reliability of morning, before-dinner, and at-bedtime home blood pressure measurements in patients with hypertension.

机构信息

Division of Cardiovascular Medicine, Department of Medicine, Jichi Medical University School of Medicine, Shimotsuke, Japan.

Higashiagatsuma-machi National Health Insurance Clinic, Gunma, Japan.

出版信息

J Clin Hypertens (Greenwich). 2018 Feb;20(2):315-323. doi: 10.1111/jch.13165. Epub 2018 Jan 6.

Abstract

The authors evaluated differences in the reliability of home blood pressure measurements taken in the morning, before dinner, and at bedtime. Forty-eight patients with hypertension (age range, 50-89 years; mean age, 76.4 years) measured their home blood pressure using a validated automatic information/communication technology-based device for 14 consecutive days. Those days were divided into the first seven days (1-7) and the following 8 to 14 days (days 8-14) and compared systolic blood pressure (SBP) reliability in the two periods for each measurement time point. In Bland-Altman analyses, morning SBP showed the least standard error of measurement (3.0 mm Hg). There were fixed biases in morning and before-dinner SBP with average limits of agreement of 3.9 and 6.4 mm Hg, respectively. For at-bedtime SBP, a random error was detected and the minimal detectable change was 13.8 mm Hg. The percentage of near-maximal variation of morning SBP was the smallest at 18.1%. Morning SBP therefore provided the most reliable home blood pressure value in the day.

摘要

作者评估了在家中清晨、饭前和睡前测量血压的可靠性差异。48 名高血压患者(年龄 50-89 岁;平均年龄 76.4 岁)使用经过验证的基于自动信息/通信技术的设备连续 14 天测量其家庭血压。这些天分为前 7 天(1-7)和接下来的 8-14 天(第 8-14 天),并比较了两个时间段内每个测量时间点的收缩压(SBP)可靠性。在 Bland-Altman 分析中,清晨 SBP 的测量标准误差最小(3.0mmHg)。清晨和饭前 SBP 存在固定偏差,平均允许误差分别为 3.9 和 6.4mmHg。对于睡前 SBP,检测到随机误差,最小可检测变化为 13.8mmHg。清晨 SBP 的最大变化百分比最小,为 18.1%。因此,清晨 SBP 提供了一天中最可靠的家庭血压值。

相似文献

引用本文的文献

本文引用的文献

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验