This study was conducted to test, in mountain running route conditions, the accuracy of the Polar V800™ monitor as a suitable device for monitoring the heart rate variability (HRV) of runners.
Eighteen healthy subjects ran a route that included a range of running slopes such as those encountered in trail and ultra-trail races. The comparative study of a V800 and a Holter SEER 12 ECG Recorder™ included the analysis of RR time series and short-term HRV analysis. A correction algorithm was designed to obtain the corrected Polar RR intervals. Six 5-min segments related to different running slopes were considered for each subject.
The correlation between corrected V800 RR intervals and Holter RR intervals was very high (r = 0.99, p < 0.001), and the bias was less than 1 ms. The limits of agreement (LoA) obtained for SDNN and RMSSD were (- 0.25 to 0.32 ms) and (- 0.90 to 1.08 ms), respectively. The effect size (ES) obtained in the time domain HRV parameters was considered small (ES < 0.2). Frequency domain HRV parameters did not differ (p > 0.05) and were well correlated (r ≥ 0.96, p < 0.001).
Narrow limits of agreement, high correlations and small effect size suggest that the Polar V800 is a valid tool for the analysis of heart rate variability in athletes while running high endurance events such as marathon, trail, and ultra-trail races.
本研究旨在测试 Polar V800™ 监测器在山地跑步路线条件下作为监测跑步者心率变异性 (HRV) 的合适设备的准确性。
18 名健康受试者在包括各种跑步坡度的路线上跑步,例如在越野和超长距离比赛中遇到的坡度。对 V800 和 Holter SEER 12 ECG Recorder™ 的比较研究包括对 RR 时间序列和短期 HRV 分析的分析。设计了一种校正算法来获得校正后的 Polar RR 间隔。每位受试者考虑了与不同跑步坡度相关的六个 5 分钟段。
校正后的 V800 RR 间隔与 Holter RR 间隔之间的相关性非常高(r=0.99,p<0.001),偏差小于 1 ms。SDNN 和 RMSSD 的一致性界限(LoA)分别为(-0.25 至 0.32 ms)和(-0.90 至 1.08 ms)。时域 HRV 参数的效应量(ES)被认为较小(ES<0.2)。频域 HRV 参数没有差异(p>0.05),并且相关性良好(r≥0.96,p<0.001)。
狭窄的一致性界限、高相关性和小的效应量表明,Polar V800 是分析马拉松、越野和超长距离比赛等高耐力运动中运动员心率变异性的有效工具。