To assess if color measurement with intraoral scanner correlates with digital colorimeter and to evaluate the possibility of application of a digital scanner for shade selection.
The Lab* values of the five shade tabs (A1, A2, A3, A3.5, and A4) were obtained with an intraoral scanner (TRIOS Pod) and a colorimeter (ShadeEye). Both devices were calibrated according to the manufacturer's instructions before measurements. Color measurement values were compared with paired t-test, and a Pearson's correlation analysis was performed to evaluate the relationship of two methods.
The Lab* values of the colorimeter were significantly different from those of the digital scanner (p < 0.001). The L* and b* values of both methods were strongly correlated with each other (both p < 0.05). The device repeatability in both methods were reported to be excellent (p < 0.05). Within the limitations of this study, color measurements with digital intraoral scanners and computer-assisted image analysis were in accordance with those of the colorimeter with respect to L* and b* values; however, all the coordinates of shade tabs were significantly different between two methods.
The digital intraoral scanner may not be used as the primary method of color selection in clinical practices, considering significant differences in color parameters with colorimeter. The scanner's capability in shade selection should be further evaluated.
评估口腔内扫描仪的颜色测量值与数字比色计是否相关,并评估数字扫描仪用于比色选择的可能性。
使用口腔内扫描仪(TRIOS Pod)和比色计(ShadeEye)获取五个比色片(A1、A2、A3、A3.5 和 A4)的 Lab*值。在测量之前,根据制造商的说明对两个设备进行了校准。使用配对 t 检验比较颜色测量值,并进行 Pearson 相关性分析以评估两种方法之间的关系。
比色计的 Lab值与数字扫描仪的 Lab值显著不同(p < 0.001)。两种方法的 L和 b值彼此之间具有很强的相关性(均 p < 0.05)。两种方法的设备重复性均被报道为优秀(p < 0.05)。在本研究的限制范围内,数字口腔内扫描仪和计算机辅助图像分析的颜色测量值与比色计的 L和 b值一致;然而,两种方法的比色片所有坐标均存在显著差异。
考虑到与比色计相比颜色参数存在显著差异,数字口腔内扫描仪可能不能用作临床比色选择的主要方法。应进一步评估扫描仪在比色选择方面的能力。