近视小切口透镜切除术与准分子激光原位角膜磨镶术术后角膜生物力学比较:一项离体研究

Comparison of corneal biomechanics after myopic small-incision lenticule extraction compared to LASIK: an ex vivo study.

作者信息

Kanellopoulos Anastasios John

机构信息

Department of Ophthalmology, Laservision Clinical and Research Institute, Athens, Greece.

Department of Ophthalmology, New York University Medical School, New York, NY, USA.

出版信息

Clin Ophthalmol. 2018 Jan 25;12:237-245. doi: 10.2147/OPTH.S153509. eCollection 2018.

Abstract

PURPOSE

To investigate ex vivo potentially different corneal biomechanical properties after small-incision lenticule extraction (SMILE) versus LASIK for myopic correction.

METHODS

Thirty human donor corneas were subjected to either myopic SMILE or femtosecond laser-assisted LASIK. Donor corneas were assigned to six investigative groups: Group A, -3.00 D (diopters) SMILE; Group B, -8.00 D SMILE; Group C, -3.00 D LASIK; and Group D, -8.00 D LASIK. Additionally, two control groups were formed: Group E, SMILE and Group F, LASIK. All groups consisted of five corneas, randomly allocated. The corneas in the control groups were subjected to the corresponding femtosecond-laser lamellar cuts but not to tissue removal. Evaluation of biomechanical tensile strength was conducted by biaxial force application. Primary outcome measures were stress at 10% and 15% strain, and Young's modulus at 10% and 15% strain.

RESULTS

In SMILE, the average relative difference (Δ) of the four outcome measures was -34.46% for -3.00 D correction versus control Group E and -49.34% for -8.00 D correction versus control Group E. In LASIK, average Δ was -24.88% for -3.00 D correction versus control, and -52.73% for -8.00 D correction versus control. All these differences were statistically significant; SMILE compared to LASIK for the same myopic correction appears to result in more biomechanical reduction for -3.00 D corrections by -26%, while a nonstatistically significant difference was noted in -8.00 D corrections.

CONCLUSION

Both SMILE and LASIK procedures do substantially alter corneal biomechanical properties, and the degree of tensile strength reduction is statistically significantly correlated to the extent of myopic correction. Additionally, SMILE procedure seems to result in more tensile strength reduction in lower myopic corrections compared to LASIK, and similar tensile strength reduction to LASIK in higher myopic corrections when compared to LASIK.

摘要

目的

研究小切口透镜切除术(SMILE)与准分子激光原位角膜磨镶术(LASIK)矫正近视后离体角膜潜在的不同生物力学特性。

方法

30只人供体角膜分别接受近视SMILE或飞秒激光辅助LASIK手术。供体角膜被分为6个研究组:A组,-3.00 D(屈光度)SMILE;B组,-8.00 D SMILE;C组,-3.00 D LASIK;D组,-8.00 D LASIK。此外,还形成了两个对照组:E组,SMILE;F组,LASIK。所有组均由5只角膜组成,随机分配。对照组的角膜接受相应的飞秒激光板层切割,但不进行组织切除。通过双轴力施加来评估生物力学拉伸强度。主要观察指标为10%和15%应变时的应力以及10%和15%应变时的杨氏模量。

结果

在SMILE中,对于-3.00 D矫正,与E组对照相比,四个观察指标的平均相对差异(Δ)为-34.46%;对于-8.00 D矫正,与E组对照相比,平均相对差异为-49.34%。在LASIK中,对于-3.00 D矫正,与对照组相比,平均Δ为-24.88%;对于-8.00 D矫正,与对照组相比,平均Δ为-52.73%。所有这些差异均具有统计学意义;对于相同的近视矫正,与LASIK相比,SMILE在-3.00 D矫正时似乎导致更多的生物力学降低,降低了26%,而在-8.00 D矫正时差异无统计学意义。

结论

SMILE和LASIK手术均会显著改变角膜生物力学特性,拉伸强度降低程度与近视矫正程度在统计学上显著相关。此外,与LASIK相比,SMILE手术在较低近视矫正时似乎导致更多的拉伸强度降低,而在较高近视矫正时与LASIK相比拉伸强度降低程度相似。