Repeatable and reproducible measurements of myocardial oxidative metabolism, blood flow and external efficiency using C-acetate PET.

机构信息

Division of Cardiology, University of Ottawa Heart Institute, 40 Ruskin St., Ottawa, ON, K1Y4W7, Canada.

出版信息

J Nucl Cardiol. 2018 Dec;25(6):1912-1925. doi: 10.1007/s12350-018-1206-y. Epub 2018 Feb 16.

Abstract

BACKGROUND

Non-invasive approaches to investigate myocardial efficiency can help track the progression of heart failure (HF). This study evaluates the repeatability and reproducibility of C-acetate positron emission tomography (PET) imaging of oxidative metabolism.

METHODS AND RESULTS

Dynamic C-acetate PET scans were performed at baseline and followup (47 ± 22 days apart) in 20 patients with stable HF with reduced ejection fraction. Two observers blinded to patients' clinical data used FlowQuant to evaluate test-retest repeatability, as well as intra- and inter-observer reproducibility of C-acetate tracer uptake and clearance rates, for the measurement of myocardial oxygen consumption (MVO), myocardial external efficiency (MEE), work metabolic index (WMI), and myocardial blood flow. Reproducibility and repeatability were evaluated using intra-class-correlation (ICC) and Bland-Altman coefficient-of-repeatability (CR). Test-retest correlations and repeatability were better for MEE and WMI compared to MVO. All intra- and inter-observer correlations were excellent (ICC = 0.95-0.99) and the reproducibility values (CR = 3%-6%) were significantly lower than the test-retest repeatability values (22%-54%, P < 0.001). Repeatability was improved for all parameters using a newer PET-computed tomography (CT) scanner compared to older PET-only instrumentation.

CONCLUSION

C-acetate PET measurements of WMI and MEE exhibited excellent test-retest repeatability and operator reproducibility. Newer PET-CT scanners may be preferred for longitudinal tracking of cardiac efficiency.

摘要

背景

非侵入性方法可用于研究心肌效率,帮助跟踪心力衰竭(HF)的进展。本研究评估了氧化代谢的 C-乙酸盐正电子发射断层扫描(PET)成像的可重复性和再现性。

方法和结果

在 20 例射血分数降低的稳定 HF 患者中,在基线和随访(相隔 47±22 天)时进行了动态 C-乙酸盐 PET 扫描。两名观察者对患者的临床数据不知情,使用 FlowQuant 评估了 C-乙酸盐示踪剂摄取和清除率的测试-重测重复性,以及观察者内和观察者间重复性,用于测量心肌耗氧量(MVO)、心肌外效率(MEE)、工作代谢指数(WMI)和心肌血流。使用组内相关系数(ICC)和 Bland-Altman 可重复性系数(CR)评估了重复性和可重复性。与 MVO 相比,MEE 和 WMI 的测试-重测相关性和可重复性更好。所有观察者内和观察者间相关性均非常好(ICC=0.95-0.99),并且再现性值(CR=3%-6%)明显低于测试-重测重复性值(22%-54%,P<0.001)。与仅使用旧的 PET 仪器相比,使用较新的 PET-CT 扫描仪可提高所有参数的可重复性。

结论

WMI 和 MEE 的 C-乙酸盐 PET 测量值具有出色的测试-重测重复性和操作员再现性。对于心脏效率的纵向跟踪,可能更喜欢使用较新的 PET-CT 扫描仪。