Gastrointest Endosc. 2018 Aug;88(2):335-344.e2. doi: 10.1016/j.gie.2018.02.043. Epub 2018 Mar 9.
Devices used to improve polyp detection during colonoscopy have seldom been compared with each other.
We performed a 3-center prospective randomized trial comparing high-definition (HD) forward-viewing colonoscopy alone to HD with Endocuff to HD with EndoRings to the full spectrum endoscopy (FUSE) system. Patients were age ≥50 years and had routine indications and intact colons. The study colonoscopists were all proven high-level detectors. The primary endpoint was adenomas per colonoscopy (APC).
Among 1188 patients who completed the study, APC with Endocuff (APC mean ± standard deviation: 1.82 ± 2.58), EndoRings (1.55 ± 2.42), and standard HD colonoscopy (1.53 ± 2.33) were all higher than FUSE (1.30 ± 1.96; P < .001 for APC). The APC for Endocuff was higher than standard HD colonoscopy (P = .014). Mean cecal insertion times with FUSE (468 ± 311 seconds) and EndoRings (403 ± 263 seconds) were both longer than with Endocuff (354 ± 216 seconds; P = .006 and .018, respectively).
For high-level detectors at colonoscopy, forward-viewing HD instruments dominate the FUSE system, indicating that for these examiners image resolution trumps angle of view. Further, Endocuff is a dominant strategy over EndoRings and no mucosal exposure device on a forward-viewing HD colonoscope. (Clinical trial registration number: NCT02345889.).
用于提高结肠镜检查中息肉检测率的设备很少被相互比较。
我们进行了一项 3 中心前瞻性随机试验,比较了单独使用高清(HD)前视结肠镜检查、HD 联合 Endocuff、HD 联合 EndoRings 与全光谱内镜(FUSE)系统。患者年龄≥50 岁,有常规适应证且结肠完整。研究结肠镜检查医师均为高水准检测者。主要终点为每例结肠镜检查的腺瘤数(APC)。
在完成研究的 1188 例患者中,Endocuff(APC 平均值±标准差:1.82±2.58)、EndoRings(1.55±2.42)和标准 HD 结肠镜检查(1.53±2.33)的 APC 均高于 FUSE(1.30±1.96;APC 差异有统计学意义,P<0.001)。Endocuff 的 APC 高于标准 HD 结肠镜检查(P=0.014)。FUSE(468±311 秒)和 EndoRings(403±263 秒)的盲肠插入中位时间均长于 Endocuff(354±216 秒;P=0.006 和 0.018)。
对于结肠镜检查的高水准检测者,前视 HD 仪器优于 FUSE 系统,这表明对于这些检查者,图像分辨率优于视野角度。此外,Endocuff 优于 EndoRings,在 HD 前视结肠镜检查中没有黏膜暴露装置。(临床试验注册号:NCT02345889.)。