评估各种参数对调强放疗验证中二维二极管阵列伽马指数值的影响。

Evaluating the Impact of Various Parameters on the Gamma Index Values of 2D Diode Array in IMRT Verification.

作者信息

Jabbari Keyvan, Pashaei Fakhereh, Ay Mohammad R, Amouheidari Alireza, Tavakoli Mohammad B

机构信息

Department of Medical Physics, Isfahan University of Medical Sciences, Isfahan, Iran.

Department of Medical Physics and Biomedical Engineering, Isfahan Milad Hospital, Isfahan, Iran.

出版信息

J Med Signals Sens. 2018 Jan-Mar;8(1):31-38. doi: 10.4103/jmss.JMSS_15_17.

Abstract

BACKGROUND

MapCHECK2 is a two-dimensional diode arrays planar dosimetry verification system. Dosimetric results are evaluated with gamma index. This study aims to provide comprehensive information on the impact of various factors on the gamma index values of MapCHECK2, which is mostly used for IMRT dose verification.

METHODS

Seven fields were planned for 6 and 18 MV photons. The azimuthal angle is defined as any rotation of collimators or the MapCHECK2 around the central axis, which was varied from 5 to -5°. The gantry angle was changed from -8 to 8°. Isodose sampling resolution was studied in the range of 0.5 to 4 mm. The effects of additional buildup on gamma index in three cases were also assessed. Gamma test acceptance criteria were 3%/3 mm.

RESULTS

The change of azimuthal angle in 5° interval reduced gamma index value by about 9%. The results of putting buildups of various thicknesses on the MapCHECK2 surface showed that gamma index was generally improved in thicker buildup, especially for 18 MV. Changing the sampling resolution from 4 to 2 mm resulted in an increase in gamma index by about 3.7%. The deviation of the gantry in 8° intervals in either directions changed the gamma index only by about 1.6% for 6 MV and 2.1% for 18 MV.

CONCLUSION

Among the studied parameters, the azimuthal angle is one of the most effective factors on gamma index value. The gantry angle deviation and sampling resolution are less effective on gamma index value reduction.

摘要

背景

MapCHECK2是一种二维二极管阵列平面剂量学验证系统。剂量学结果通过伽马指数进行评估。本研究旨在提供关于各种因素对MapCHECK2伽马指数值影响的全面信息,MapCHECK2主要用于调强放疗剂量验证。

方法

针对6和18兆伏光子计划了七个射野。方位角定义为准直器或MapCHECK2围绕中心轴的任何旋转,其范围从5°到 -5°。机架角度从 -8°改变到8°。在0.5至4毫米范围内研究了等剂量采样分辨率。还评估了三种情况下额外建成对伽马指数的影响。伽马测试验收标准为3%/3毫米。

结果

以5°间隔改变方位角使伽马指数值降低约9%。在MapCHECK2表面放置不同厚度建成物的结果表明,较厚的建成物通常会改善伽马指数,尤其是对于18兆伏的情况。将采样分辨率从4毫米改为2毫米导致伽马指数增加约3.7%。机架在任一方向上以8°间隔的偏差对6兆伏的伽马指数仅改变约1.6%,对18兆伏的伽马指数改变约2.1%。

结论

在所研究的参数中,方位角是对伽马指数值影响最有效的因素之一。机架角度偏差和采样分辨率对伽马指数值降低的影响较小。