Department of Periodontics and Implantology, Stomatological Hospital, Southern Medical University, Guangzhou, 510280, China.
J Mater Sci Mater Med. 2018 Mar 30;29(4):43. doi: 10.1007/s10856-018-6049-1.
This study was designed to compare the early osseointegration of titanium surfaces prepared via laser-treated/acid-etched (LA) and sandblasted/acid-etched (SLA) in dogs. Titanium implants were divided into two groups: Surfaces of the experimental group were treated via LA, while in the control group, surfaces were treated via SLA. The physical and chemical properties of LA and SLA surfaces were tested and compared. Sixteen implants with LA or SLA surfaces were placed into the tibias of four beagle dogs, each treatment group received two implants per single tibia. The dogs were sacrificed two and four weeks after implant placement. Scanning electron microscopy showed that both the LA and SLAs surface exhibited rough structures with micro pores sized 1-3 μm. In the LA surface, regular melting points were observed. However, in the SLA surface, the structure was irregular and few oxide aluminum particles still remained. Only titanium and a small amount of titanium compounds were detected on LA surfaces, while Al was found of SLA surfaces. The LA surface roughness was above that of SLA surfaces (LA: Ra: 2.1 μm; SLA: Ra :1.53 μm; P < 0.01). Both groups exhibited good osseointegration and no significant differences were found in the BIC% at two or four weeks between both groups (P > 0.05). Both groups exhibited good osseointegration; however, the LA surface was cleaner and more uniform than the SLA surface, and no significant differences were found between both groups.
本研究旨在比较经激光处理/酸蚀(LA)和喷砂/酸蚀(SLA)处理的钛表面的早期骨整合情况。将钛植入物分为两组:实验组的表面经 LA 处理,而对照组的表面经 SLA 处理。测试并比较了 LA 和 SLA 表面的物理和化学性质。将具有 LA 或 SLA 表面的 16 个植入物分别植入 4 只比格犬的胫骨中,每个治疗组的每只胫骨植入 2 个植入物。植入物放置 2 周和 4 周后,处死犬。扫描电子显微镜显示,LA 和 SLA 表面均呈现出具有 1-3μm 微孔的粗糙结构。在 LA 表面,观察到规则的熔融点。然而,在 SLA 表面,结构不规则,仍有少量氧化铝颗粒存在。仅在 LA 表面检测到钛和少量钛化合物,而在 SLA 表面检测到 Al。LA 表面粗糙度高于 SLA 表面(LA:Ra:2.1μm;SLA:Ra:1.53μm;P<0.01)。两组均表现出良好的骨整合,两组在 2 周和 4 周时的 BIC% 之间无显著差异(P>0.05)。两组均表现出良好的骨整合;然而,LA 表面比 SLA 表面更清洁、更均匀,两组之间无显著差异。