Oikonomidis I L, Tsouloufi T K, Soubasis N, Kritsepi-Konstantinou M
Diagnostic Laboratory, School of Veterinary Medicine, Faculty of Health Sciences, Aristotle University of Thessaloniki, Thessaloniki, Greece.
Diagnostic Laboratory, School of Veterinary Medicine, Faculty of Health Sciences, Aristotle University of Thessaloniki, Thessaloniki, Greece.
Vet J. 2018 Apr;234:48-54. doi: 10.1016/j.tvjl.2018.02.005. Epub 2018 Feb 14.
Haemoglobin A1c (HbA1c), the major fraction of glycated haemoglobin, is widely used for the diagnosis and monitoring of diabetes mellitus in human beings. However, there is a paucity of literature on the most reliable methods available for measurement of canine HbA1c. The aim of this study was to validate a new automated capillary electrophoresis assay for canine HbA1c, to generate a reference interval and to assess the overlap performance of the assay for the diagnosis of diabetes mellitus. Eighty-three blood samples treated with EDTA were included in the study, comprising 63 from healthy dogs and 20 from diabetic dogs. Linearity was assessed by mixing canine samples of known HbA1c percentage in different proportions, precision was assessed by repeated (n=8) measurement of five canine samples, and stability was assessed by measuring canine samples stored at 4°C for 96h and -20°C for 4 weeks. The robust method was used to determine the reference interval. The assay was demonstrated to be linear (R=0.943). Intra-assay and inter-assay coefficients of variation (CVs) were 4.8% and 7.0%, respectively. CVs for blood samples stored at 4°C and -20°C were 7.2% and 11.2%, respectively. The reference interval was 0.6-2.7%. Dogs with diabetes mellitus had significantly (P<0.001) higher mean HbA1c (5.24±0.88%) compared to the reference population (1.64±0.55%), with no overlap between results. A HbA1c cut-off of 3.3% clearly differentiated diabetic from healthy dogs. The capillary electrophoresis assay was properly validated for canine HbA1c and the reference interval was determined, while the overlap performance of the assay was excellent.
糖化血红蛋白A1c(HbA1c)是糖化血红蛋白的主要成分,广泛用于人类糖尿病的诊断和监测。然而,关于测量犬类HbA1c最可靠方法的文献却很少。本研究的目的是验证一种用于犬类HbA1c的新型自动毛细管电泳测定法,生成参考区间,并评估该测定法在糖尿病诊断中的重叠性能。本研究纳入了83份用乙二胺四乙酸(EDTA)处理的血样,其中63份来自健康犬,20份来自糖尿病犬。通过以不同比例混合已知HbA1c百分比的犬类样本评估线性,通过对五个犬类样本进行重复(n = 8)测量评估精密度,通过测量在4°C下储存96小时和在-20°C下储存4周的犬类样本评估稳定性。采用稳健方法确定参考区间。该测定法被证明具有线性(R = 0.943)。批内和批间变异系数(CV)分别为4.8%和7.0%。在4°C和-20°C下储存的血样的CV分别为7.2%和11.2%。参考区间为0.6 - 2.7%。与参考群体(1.64±0.55%)相比,糖尿病犬的平均HbA1c显著更高(P<0.001,5.24±0.88%),结果之间没有重叠。HbA1c临界值为3.3%可明显区分糖尿病犬和健康犬。毛细管电泳测定法已针对犬类HbA1c进行了适当验证,并确定了参考区间,同时该测定法的重叠性能优异。