Messaoud Mouna, Marsiquet Cyril, Revol-Cavalier Frédéric, Rat Venceslass, Marchand Gilles
Assistant Professor, Laboratory of Industrial Chemistry (LCI), Soukra avenue Km 3.5, Sfax, Tunisia.
Research Engineer, The French Alternative Energies and Atomic Energy Commission (CEA), MINATEC Campus, F-38054 Grenoble Cedex.
J Wound Care. 2018 Jun 2;27(6):385-391. doi: 10.12968/jowc.2018.27.6.385.
Few studies have investigated methods to monitor the moisture distribution and filling percentage of dressings during wound management. In this study, a new system allowing moisture monitoring across the wound dressing to be measured is examined.
The system is composed of a wound bed model with a fluid injection system to mimic exudate flow, a flexible sensor array and data acquisition software. The sensor is composed of 14 flexible electrode arrays, screen-printed on a flexible support and placed on the top of a wound dressing. The system is used to evaluate the performance of a foam-based dressing model.
During constant injection of fluid, the wound dressing absorbed moisture at the wound interface throughout the experiment, and expanded as the fluid spread from the wound bed to the edging areas of the foam. The in-time monitoring by the use of the screen-printed electrodes allowed a mapping of the dressing wet surface and estimation of the foam saturation (filling percentage) based on a simple acquisition method without the need to remove the dressing from the wound bed.
The findings of this study propose a non-invasive method to monitor the filling of the wound dressing and consequently, a potential solution for determining the optimal dressing change during wound management without causing irritation or further damage to the periwound skin.
很少有研究探讨在伤口处理过程中监测敷料水分分布和填充率的方法。在本研究中,对一种能够测量伤口敷料水分分布的新系统进行了检测。
该系统由一个带有模拟渗出液流动的流体注射系统的伤口床模型、一个柔性传感器阵列和数据采集软件组成。传感器由14个柔性电极阵列组成,通过丝网印刷在柔性支撑物上,并放置在伤口敷料顶部。该系统用于评估基于泡沫的敷料模型的性能。
在持续注入液体的过程中,整个实验期间伤口敷料在伤口界面处吸收水分,并随着液体从伤口床扩散到泡沫边缘区域而膨胀。通过使用丝网印刷电极进行实时监测,可以绘制敷料湿润表面的图谱,并基于一种简单的采集方法估计泡沫饱和度(填充率),而无需将敷料从伤口床上取下。
本研究结果提出了一种无创监测伤口敷料填充情况的方法,从而为在伤口处理过程中确定最佳换药时间提供了一种潜在的解决方案,且不会对伤口周围皮肤造成刺激或进一步损伤。