Evidence on radiation dose reduction using rectangular collimation: a systematic review.

机构信息

School of Dentistry, Faculty of Medicine and Dentistry, University of Alberta, Edmonoton, AB, Canada.

出版信息

Int Dent J. 2019 Apr;69(2):84-97. doi: 10.1111/idj.12411. Epub 2018 Jun 29.

Abstract

BACKGROUND

The goal of an intraoral radiograph is to be a valuable diagnostic tool while keeping the radiation exposure dose as low as reasonably achievable. The International Commission on Radiological Protection (ICRP) has strong recommendations for patient dose-reduction strategies.

OBJECTIVE

To answer the following research question: 'In either patient or in phantoms, does rectangular collimation compared with other forms of collimation reduce the radiation absorbed dose during intraoral imaging?'

METHODS

The search strategy was developed specifically for each electronic database. Information was collected from MEDLINE, all EBM Reviews including the Cochrane Library, EMBASE, LILACS, PubMed and Web of Science, and from a search of the grey literature via Google Scholar. Studies were included if they reported the radiation dose in intraoral imaging using rectangular collimation compared with any other type of collimation or the absence of it. Studies that did not involve comparison/control groups were excluded. A qualitative appraisal of the included studies was performed. The Methodological Index for NOn-Randomised Studies (MINORS) tool assessed the potential risk of bias among the studies, while the Grading of Recommendations, Assessment, Development and Evaluation (GRADE) approach determined the level of available evidence.

RESULTS

Thirteen studies using different types of receptors, including D-speed film, E-speed film, phosphor storage plates (PSP) and direct digital sensors, were included. Different methods to access absorbed and effective dose by exploring distinct technical parameters, such as distance object-receptor, long- and short-cone, kVp and mA, exposure time and beam filtering, were used. High heterogeneity between the studies was found. The level of evidence was classified as moderate. The radiation dose reduction ranged from 40% to 92% compared with circular collimation. The limitations of the cross-sectional design and in vivo studies were discussed. As this systematic review applied validated tools and guidelines designed for population-based studies, the application for studies based on computer simulations and dosimetry measurements was challenging.

CONCLUSION

There is moderate evidence suggesting that rectangular collimation significantly reduces radiation dose when compared with round collimation, justifying its implementation in clinical settings/private practices while taking intraoral radiographs.

摘要

背景

口腔内射线照相的目标是成为有价值的诊断工具,同时将辐射暴露剂量保持在尽可能低的合理可达到的水平。国际辐射防护委员会(ICRP)对患者剂量降低策略有强烈的建议。

目的

回答以下研究问题:“在患者或体模中,与其他准直形式相比,矩形准直是否会减少口腔内成像期间的辐射吸收剂量?”

方法

针对每个电子数据库专门制定了搜索策略。信息来自 MEDLINE、所有 EBM 综述,包括 Cochrane 图书馆、EMBASE、LILACS、PubMed 和 Web of Science,以及通过 Google Scholar 对灰色文献的搜索。如果研究报告了使用矩形准直与任何其他类型的准直或没有准直相比的口腔内成像的辐射剂量,则将其纳入研究。未涉及对照/对照组的研究被排除在外。对纳入的研究进行了定性评估。非随机研究方法学指数(MINORS)工具评估了研究中的潜在偏倚风险,而推荐评估、发展和评估(GRADE)方法确定了现有证据的水平。

结果

共纳入了 13 项使用不同类型接收器的研究,包括 D 速度胶片、E 速度胶片、磷存储板(PSP)和直接数字传感器。使用不同的方法通过探索不同的技术参数来获取吸收剂量和有效剂量,例如距离物体-接收器、长和短圆锥体、kVp 和 mA、曝光时间和束滤波。研究之间存在高度异质性。证据水平被归类为中等。与圆形准直相比,辐射剂量降低幅度为 40%至 92%。讨论了横截面设计和体内研究的局限性。由于本系统评价应用了针对基于人群的研究设计的经过验证的工具和指南,因此对于基于计算机模拟和剂量测量的研究的应用具有挑战性。

结论

有中等证据表明,与圆形准直相比,矩形准直可显著降低辐射剂量,这证明了在临床环境/私人诊所中实施矩形准直的合理性,同时进行口腔内射线照相。

相似文献

7
The value of thyroid shielding in intraoral radiography.口腔内放射摄影中甲状腺防护的价值。
Dentomaxillofac Radiol. 2016;45(5):20150407. doi: 10.1259/dmfr.20150407. Epub 2016 Mar 23.

引用本文的文献

4
Enhancing intraoral radiographic technique: introducing "Nallan's Lines".提高口腔内放射摄影技术:引入“纳兰线”。
Front Oral Health. 2024 Dec 20;5:1498117. doi: 10.3389/froh.2024.1498117. eCollection 2024.
5
Evaluation of X-ray Beam Collimation in Adult Chest Radiography.成人胸部X线摄影中X线束准直的评估
Cureus. 2024 Sep 2;16(9):e68489. doi: 10.7759/cureus.68489. eCollection 2024 Sep.

本文引用的文献

6
NCRP report 160 and what it means for medical imaging and nuclear medicine.NCRP报告160及其对医学成像和核医学的意义。
J Nucl Med Technol. 2013 Dec;41(4):255-60. doi: 10.2967/jnmt.113.128728. Epub 2013 Oct 31.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验