In Vitro Effect of Mastication on the Retention and Wear of Locator Attachments in a Flat Mandibular Ridge Model.

机构信息

Department of Prosthodontics, Lebanese University, Beirut, Lebanon.

Advanced Education Program in Prosthodontics, Loma Linda University, Loma Linda, CA.

出版信息

J Prosthodont. 2019 Feb;28(2):e744-e751. doi: 10.1111/jopr.12940. Epub 2018 Jul 2.

Abstract

PURPOSE

The effect of masticatory loads on the retention of overdenture attachments is poorly documented. The aim of this in vitro study is to assess the effect of simulated mastication on the retentive properties and dimensions of Locator inserts.

MATERIALS AND METHODS

30 specimens simulating nonanatomic edentulous flat ridges with two implant replicas each were fabricated. Overdenture units were connected to the implants with Locator attachments and 3 types of inserts: transparent (Group T; n = 10 pairs), pink (Group P; n = 10 pairs), and blue (Group B; n = 10 pairs). The overdenture units were subjected to simulated bi-axial masticatory loads of 68.6 N. Locator retention was assessed using axial dislodging forces at baseline (T0) and following 100,000 (T1) and 200,000 (T2) masticatory cycles. The inner diameter of the insert (XY) and the diameter of the central core (AB) were measured under stereomicroscope at T0 and T2. Retention changes and dimensional variations of the inserts were statistically analyzed.

RESULTS

The 3 groups showed significantly different retentions with the highest values recorded for group T followed by group P, and finally group B at T0, T1, and T2. Groups T and P were not affected by loading while group B showed a significant mean retention loss from T0 to T1. XY and AB were significantly different between the 3 color-coded inserts at baseline and at T2. No correlation could be established between retention changes and dimensional variations of the 3 types of inserts.

CONCLUSIONS

Within the limitations of this in vitro study, simulated mastication seems to significantly affect the extra-light blue Locator inserts but not the more-retentive ones. The transparent and pink inserts may require less frequent replacements and could therefore be recommended under clinical conditions.

摘要

目的

咀嚼负荷对覆盖义齿附着体固位力的影响尚未得到充分证实。本体外研究旨在评估模拟咀嚼对Locator 附着体保留性能和尺寸的影响。

材料和方法

制作 30 个模拟非解剖无牙牙槽嵴的试件,每个试件带有两个种植体模型。用 Locator 附着体将覆盖义齿单元连接到种植体上,共 3 种附着体:透明(组 T;n=10 对)、粉红色(组 P;n=10 对)和蓝色(组 B;n=10 对)。对覆盖义齿单元施加 68.6 N 的双轴咀嚼负荷。在基线(T0)以及 10 万次(T1)和 20 万次(T2)咀嚼循环后,使用轴向脱位力评估 Locator 保留情况。在体视显微镜下测量 T0 和 T2 时的附着体内部直径(XY)和中心芯直径(AB)。使用方差分析对附着体的保留变化和尺寸变化进行统计学分析。

结果

3 组的保留力有显著差异,组 T 的保留力最高,其次是组 P,最后是组 B,在 T0、T1 和 T2 时均如此。组 T 和组 P 不受加载影响,而组 B 从 T0 到 T1 时保留力显著下降。在基线和 T2 时,3 种颜色编码的附着体的 XY 和 AB 有显著差异。在 3 种附着体之间,保留力变化与尺寸变化之间没有相关性。

结论

在本体外研究的限制范围内,模拟咀嚼似乎显著影响超轻蓝色 Locator 附着体,但不影响保留力更强的附着体。透明和粉红色附着体可能需要较少的更换,因此在临床条件下可能更推荐使用。