Suppr超能文献

使用剪切波弹性成像技术对健康志愿者正常伸肌总腱进行观察者间变异性和硬度测量。

Interobserver variability and stiffness measurements of normal common extensor tendon in healthy volunteers using shear wave elastography.

作者信息

Şendur Halit Nahit, Cindil Emetullah, Cerit Mahinur, Demir Nur Betül, Şendur Aylin Billur, Oktar Suna Özhan

机构信息

Department of Radiology, Gazi University Faculty of Medicine, 06500, Beşevler, Ankara, Turkey.

Lösante Children's and Adult Hospital, Ankara, Turkey.

出版信息

Skeletal Radiol. 2019 Jan;48(1):137-141. doi: 10.1007/s00256-018-3021-6. Epub 2018 Jul 13.

Abstract

OBJECTIVES

The purposes of our study are to determine the quantitative elasticity values of normal common extensor tendon (CET) and to assess the interobserver variability of stiffness measurements using shear wave elastography (SWE).

MATERIALS AND METHODS

A total of 60 CETs of 30 (15 female, 15 male, mean age 30.2 years) healthy volunteers without any symptoms of lateral epicondylitis were examined by two radiologists. Age, sex, height, weight, body mass index (BMI), and dominant hand of all participants were noted. The first observer performed B-mode and SWE imaging, and the second observer performed only SWE imaging. Tendon thickness and stiffness values in kPa were measured.

RESULTS

The mean thickness of CETs was 3.57 ± 0.36 mm. The mean stiffness values of CETs for two observers were 45.28 ± 9.82 kPa and 45.80 ± 9.72 kPa respectively. Tendon thickness had a weak correlation with weight (r = 0.281, p = 0.03), and moderate correlation with stiffness values (r = 0.429, p < 0.001). The mean interobserver difference of CET stiffness measurements was -0.5% of the mean CET stiffness values. Range of measurement error, defined as 95% limits of agreement, was ±23.5%. There was no significant difference between absolute values of interobserver measurements (p = 0.741).

CONCLUSION

Shear wave elastography is a reproducible imaging technique for the evaluation of CET elasticity and the standard stiffness values of normal CET can be used as reference data to differentiate normal from pathological tissues.

摘要

目的

本研究的目的是确定正常伸肌总腱(CET)的定量弹性值,并评估使用剪切波弹性成像(SWE)进行硬度测量时观察者间的变异性。

材料与方法

两名放射科医生对30名(15名女性,15名男性,平均年龄30.2岁)无任何外侧上髁炎症状的健康志愿者的60条CET进行了检查。记录了所有参与者的年龄、性别、身高、体重、体重指数(BMI)和优势手。第一位观察者进行B模式和SWE成像,第二位观察者仅进行SWE成像。测量了肌腱厚度和以kPa为单位的硬度值。

结果

CET的平均厚度为3.57±0.36mm。两名观察者测得的CET平均硬度值分别为45.28±9.82kPa和45.80±9.72kPa。肌腱厚度与体重呈弱相关性(r = 0.281,p = 0.03),与硬度值呈中度相关性(r = 0.429,p < 0.001)。CET硬度测量的观察者间平均差异为CET平均硬度值的-0.5%。测量误差范围(定义为一致性界限的95%)为±23.5%。观察者间测量的绝对值之间无显著差异(p = 0.741)。

结论

剪切波弹性成像是一种可重复的成像技术,用于评估CET弹性,正常CET的标准硬度值可作为区分正常组织与病理组织的参考数据。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验