To assess sound levels of 4 high-frequency neonatal ventilators to determine whether there is a safety benefit in using modern high-frequency ventilators compared with older models.
We performed a bench study comparing noise production of the Sensormedics 3100A Oscillator, Bunnell Life Pulse Jet Ventilators Model 203 and Model 204, and Dräger VN500 in high-frequency mode. A wide range of ventilation settings was examined. All measurements were performed in triplicate using a high-fidelity sound meter, with data analyzed using ANOVA and regression analyses.
The Dräger ventilator was quietest overall, with average sound levels of 49.8 ± 0.49 dB across all settings. The average noise from the Sensormedics was 53.6 ± 2.01 dB, for Bunnell Model 203 was 54.1 ± 1.09 dB, and for Bunnell Model 204 was 53.7 ± 1.45 dB. Adjustments made to frequency/rate and mean airway pressure/positive end-expiratory pressure had minimal effect on noise, and increasing amplitude/peak inspiratory pressure resulted in significantly more noise by all ventilators. At all settings, the Sensormedics and Bunnell ventilators were louder than the Dräger, and the difference became greater as amplitude/peak inspiratory pressure was increased.
The Dräger VN500 in high-frequency mode produces significantly less noise that both the Sensormedics and Bunnell ventilators. These data suggest that using the Dräger VN500 as a high-frequency ventilator may reduce the potential for adverse outcomes created by ventilator noise.
评估 4 种高频新生儿呼吸机的声级,以确定与旧型号相比,使用现代高频呼吸机是否具有安全性优势。
我们进行了一项台架研究,比较了 Sensormedics 3100A 振荡器、Bunnell Life Pulse Jet 呼吸机 Model 203 和 Model 204 以及 Dräger VN500 在高频模式下的噪声产生情况。检查了广泛的通气设置。所有测量均使用高保真度声级计重复进行三次,使用 ANOVA 和回归分析对数据进行分析。
总体而言,Drager 呼吸机的噪音最小,所有设置的平均声级为 49.8±0.49dB。Sensormedics 的平均噪声为 53.6±2.01dB,Bunnell Model 203 为 54.1±1.09dB,Bunnell Model 204 为 53.7±1.45dB。频率/速率和平均气道压/呼气末正压的调整对噪声的影响很小,而增加幅度/峰吸气压会导致所有呼吸机产生更大的噪声。在所有设置下,Sensormedics 和 Bunnell 呼吸机的噪音都大于 Dräger,随着幅度/峰吸气压的增加,差异变得更大。
高频模式下的 Dräger VN500 产生的噪声明显小于 Sensormedics 和 Bunnell 呼吸机。这些数据表明,使用 Dräger VN500 作为高频呼吸机可能会降低呼吸机噪声引起不良后果的可能性。